Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Comparison of corneal pachymetry evolution after accelerated corneal crosslinking in keratoconus eyes using anterior
Antonia Lang1, Berthold Seitz2, Cristian Munteanu2
1Department of Ophthalmology, Saarland University Medical Center, Kirrberger Straße, Building 22, 66424, Homburg, Germany. anla00014@stud.unisaarland.de.
Purpose:
This retrospective clinical study compared longitudinal corneal pachymetry measurement evolution after accelerated corneal crosslinking (A-CXL, 9 mW/cm2, 10 min) in keratoconus (KC) eyes as measured by Pentacam HR Scheimpflug imaging (Oculus, Germany) and Casia 2 (Tomey, Japan) anterior segment optical coherence tomography (AS-OCT).
Methods:
218 eyes covering the full topographic KC (TKC) severity spectrum that underwent A-CXL between 2017 and 2021 were examined at five time points (before, 6 weeks, 6 months, 1 year after A-CXL and at the latest follow-up) using both Pentacam HR and Casia 2. Central (CCT) and thinnest (TCT) corneal thickness were assessed. Corneal densitometry and TKC classification were assessed by Pentacam HR only. Intraoperative CCT evolution was monitored using ultrasound (Pachymeter SP-3000, Tomey, Japan) to ensure that CCT did not fall below 400 µm.
Results:
CCT and TCT measured by Pentacam HR were significantly lower than preoperatively during the first postoperative year, while Casia 2 CCT and TCT measurements were not significantly lower than preoperatively at any time. The highest agreement of Pentacam HR and Casia 2 measurements occurred six weeks after A-CXL. Corneal densitometry peaked six weeks after A-CXL and decreased thereafter without reaching pre-A-CXL values. Ultrasound revealed significant CCT thinning during A-CXL.
Conclusion:
Corneal Scheimpflug and AS-OCT based thickness measurements are not interchangeable. Corneal thinning after A-CXL detected by Pentacam HR was not observed in Casia 2 measurements. Since higher densitometry values did not correlate with a greater inter-device measurement difference, Scheimpflug-based corneal thickness measurements should be interpreted cautiously during the first year after A-CXL.
More Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...