Radiation Dose Alert: CT Overexposure As a Result of Patient Off-Centering
1Epicura Hospital, Belgium.
Abstract:
A high cumulative radiation dose (3394 mGy.cm, 51 mSv) delivered by two acquisitions on the abdomen and pelvis and corresponding to more than 10 times the local median dose for this examination was retrospectively detected thanks to the Dose Archiving and Communicating System (DACS). Retrospective analysis of the examination showed that the cause of this high dose was an off-centering of the patient-the table being in a too high position-disabling a correct diagnosis of an umbilical herniation that was visible on a repeated acquisition. Interestingly, the CTDIvol of the second acquisition was 40% lower than that of the first acquisition, a well-known effect of off-centering on automatic exposure control systems.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Biological Effects of Radiation
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography