Performance characterization of light-emitting diode array-based photoacoustic computed tomography (LED-PACT).
Avishek Das1, Manojit Pramanik1
1Department of Electrical and Computer Engineering, Iowa State University, Ames, Iowa 50011, USA.
Biomedical Optics Express
|December 10, 2025
Summary
A new, cost-effective light-emitting diode photoacoustic computed tomography (LED-PACT) system offers comparable imaging performance to laser-based systems. This innovation promises portable photoacoustic imaging for broader clinical use.
Area of Science:
- Biomedical Optics
- Medical Imaging
- Acoustic Imaging
Background:
- Photoacoustic computed tomography (PACT) typically relies on expensive pulsed laser diodes (PLDs).
- There is a need for more cost-effective and portable PACT systems for wider clinical adoption.
- Developing alternative light sources for PACT is crucial for advancing the field.
Purpose of the Study:
- To develop and evaluate a novel, cost-effective light-emitting diode-based PACT (LED-PACT) system.
- To compare the imaging performance of the LED-PACT system against traditional PLD-based PACT.
- To assess the feasibility of using a custom nanosecond pulsed LED array for PACT applications.
Main Methods:
- Integration of a 37-LED custom nanosecond pulsed LED array (850 nm, 100 ns pulse width) with a single-element ultrasound transducer for circular scanning.
- Driving the LED array using a homemade nanosecond pulsed current source (NSPCS) driver.
- Performance evaluation using phantom and chicken tissue studies, comparing imaging depth, resolution, and signal-to-noise ratio (SNR) against PLD-PACT.
Main Results:
- The LED-PACT system achieved a SNR of 22.4 dB (2.25 MHz UST) and a spatial resolution of 178 µm (5 MHz UST).
- A practical imaging depth of 20 mm was demonstrated in chicken tissue.
- The system operated at significantly lower optical energies (0.18 mJ) compared to PLD-PACT (1.88 mJ) while maintaining competitive performance.
- The total system cost was less than $1,000.
Conclusions:
- The developed LED-PACT system provides a cost-effective alternative to PLD-based PACT.
- The system demonstrates competitive imaging performance, including depth, resolution, and SNR.
- This LED-PACT technology holds significant potential for developing portable and widely accessible photoacoustic imaging devices for clinical translation.
Related Concept Videos
Computed Tomography
7.9K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
7.9K
Imaging Studies III: Computed Tomography
257
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
257