Detection of the Linac Jaw face angle misalignment using high energy electron beam symmetry
Song Gao1, Andres Cibrian1, Jared Ohrt1
1Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
Purpose:
To demonstrate the usage of a high energy electron beam symmetry measured at different collimator angles to detect jaws face angle misalignments on Varian TrueBeam linear accelerators (linac) during acceptance and commissioning processes.
Methods:
During the linac acceptance, all beams were steered to a symmetric shape with the gantry and collimator at 0°. We noted that the symmetries of higher energies electrons (e.g., 16 MeV, 20 MeV) changed over 2% when measured with the collimator at 90° and/or 270° for two linacs commissioned within 5-months of each other. Misaligned applicators were ruled out as the cause of this issue. Eventually, this was traced to the X-jaw face angle not matching beam divergence. After the problem was identified and the jaws were realigned by manufacture's engineers and all beams were re-steered to a symmetric shape at collimator 0° and verified at collimator 90° and 270°.
Results:
Prior to the manufacturer adjusting the jaw face angles, the maximum observed electron beam symmetry changes with 90° collimator rotation were 2.0% and 2.6% for linac 1 and linac 2, respectively. After realigning, the jaw face angles the maximum observed electron beam symmetry changes with 90° collimator rotation were within 1.0% for all beams on both linacs.
Conclusion:
We have demonstrated that changes in symmetry of high energy electron beam with 90° collimator rotation can signal that the jaw face angle is not properly aligned with beam divergence.
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal...
Eccentric Axial Loading in a Plane of Symmetry
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...