Clinical Tolerance to Induced Astigmatism in Eyes Implanted with a Single Extended Depth of Focus Intraocular Lens
Wilson Takashi Hida1, Bernardo Kaplan Moscovici2,3,4, Paulo Eduardo Souto Castro Miziara1
1Department of Ophthalmology, Hospital Oftalmológico de Brasília, Brasília, DF, Brazil.
Purpose:
To evaluate clinical tolerance to induced astigmatism in pseudophakic eyes bilaterally implanted with an extended depth of focus (EDOF) intraocular lens (AT LARA®, Carl Zeiss Meditec). We hypothesized that small amounts of induced astigmatism may be clinically well tolerated by patients implanted with this EDOF IOL.
Methods:
Prospective study including 10 patients (20 eyes) bilaterally implanted with the same EDOF IOL. Monocular defocus curves were obtained under photopic conditions after introducing cylindrical lenses of +1.00 D, +2.00 D, and +3.00 D at with-the-rule (WTR) and against-the-rule (ATR) orientations.
Results:
The IOL demonstrated good tolerance up to +1.00 D of induced astigmatism, with significant reductions in visual acuity at higher magnitudes of astigmatism. Tolerance was higher for ATR astigmatism compared to WTR, consistent with axis-dependent depth-of-focus effects.
Conclusion:
The AT LARA® EDOF IOL maintained functional visual performance with up to +1.00 D of regular astigmatism, while higher magnitudes led to clinically meaningful degradation. Axis orientation modulated performance, with ATR preserving intermediate and near vision and WTR favoring distance.