Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay
John W Pickering1,2, Stephen du Toit3, Stacey Cheer3
1Department of Emergency Medicine, Christchurch Hospital, Christchurch, New Zealand.
This study verified the QuidelOrtho TriageTrue high-sensitivity troponin I assay. The point-of-care assay demonstrated acceptable analytical performance and concordance with other methods for clinical use.
Area of Science:
- Clinical Chemistry
- Point-of-Care Testing
- Cardiovascular Diagnostics
Background:
- The QuidelOrtho TriageTrue high-sensitivity troponin I (hs-TnI) assay is a validated point-of-care (POC) diagnostic tool.
- Independent verification of its analytical performance and concordance with other assays was necessary.
Purpose of the Study:
- To independently verify the analytical performance of the TriageTrue hs-TnI assay.
- To assess the assay's concordance with other hs-TnI assays and secondary POC devices.
Main Methods:
- Precision studies (intra-assay, inter-analyzer, inter-lot) were conducted using whole blood samples.
- Interference from hemolysis was evaluated.
- Concordance was assessed using Pearson correlation and kappa statistics against other POC and laboratory analyzers.
- Comparison with secondary POC devices was performed against prespecified difference limits.
Main Results:
- Intra-assay precision CVs ranged from 2.8% to 25% across the measuring interval.
- Minimal interference was observed from hemolysis up to 10 g/L.
- High correlation coefficients (0.89-0.99) were found with most assays, except for Roche hs-TnT (0.79).
- Comparisons with 8 secondary analyzers showed results within acceptable limits.
Conclusions:
- The TriageTrue hs-TnI assay demonstrated performance consistent with the manufacturer's claims.
- The assay exhibits acceptable analytical characteristics for point-of-care use in medical settings.