Three-Dimensional Phase Resolved Functional Lung Magnetic Resonance Imaging
Published on: June 21, 2024
Self-Supervised Joint Reconstruction and Denoising of T2-Weighted PROPELLER MRI of the Lung at 0.55T
Jingjia Chen1,2, Haoyang Pei1,2,3, Christoph Maier1,2
1Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University Grossman School of Medicine, New York, New York, USA.
Purpose:
To improve 0.55T T2-weighted PROPELLER lung MRI by developing a self-supervised framework for joint reconstruction and denoising.
Methods:
T2-weighted 0.55T lung MRI datasets from 44 patients with prior COVID-19 infection were used. Each PROPELLER blade was split along the readout direction into two disjoint subsets: one subset for training an unrolled network, and the other for loss calculation. Following the Noise2Noise paradigm, this framework split k-space into two subsets with independent, matched noise but identical underlying signal, enabling joint reconstruction and denoising without external training references. For comparison, coil-wise Marchenko-Pastur Principal Component Analysis (MPPCA) denoising followed by parallel imaging reconstruction was performed. The reconstructed images were evaluated by two experienced chest radiologists.
Results:
The self-supervised model generated lung images with improved clarity, better delineation of parenchymal and airway structures, and maintained high fidelity in cases with available CT references. In addition, the proposed framework also enabled further reduction of scan time by reconstructing images with adequate diagnostic quality from only half the number of blades. The reader study confirmed that the proposed method outperformed MPPCA across all categories (Wilcoxon signed-rank test, p < 0.001), with moderate inter-reader agreement (weighted Cohen's kappa = 0.55; percentage of exact and within ±1 point agreement = 91%).
Conclusion:
By leveraging the intrinsic data redundancy in PROPELLER sampling and extending the Noise2Noise concept, the proposed self-supervised framework enabled simultaneous reconstruction and denoising of lung images at 0.55T to address the low-SNR challenge at low-field. It holds great potential for broad use in other low-field MRI applications.
More Related Videos
05:07Author Spotlight: Optimized Lung MRI Protocol with Computationally Efficient Reconstruction Methods
Published on: September 6, 2024
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...