Impact of radiotherapy plan complexity on dosimetric measurement verification: A technical evaluation of two O-ring
Zhengwen Shen1, Xin Yang1, Huanli Luo1
1Department of Radiation Oncology, Chongqing University Cancer Hospital, Chongqing, China.
Background And Purpose:
The measurement verification is an essential process of radiotherapy pre-treatment patient-specific quality assurance. The impact of plan complexity on measurement verification needs to be investigated.
Methods And Materials:
Sixty clinical radiotherapy plans were retrospectively analyzed using two types of O-ring detector arrays, ArcCHECK and ArcMap, to implement the measurement verification. Forty-three complexity indexes were calculated to comprehensively quantify plan characteristics. Spearman's correlation analysis statistically evaluated the correlations between complexity indexes and gamma passing rates (GPRs).
Results:
ArcMap achieved higher GPRs compared with ArcCHECK. Higher correlation coefficients |r| were achieved with the criteria of gamma analysis were tightened. At the 2 %/1 mm criteria, complexity indexes characterizing leaf motion, notably A4-5 (proportion of leaf accelerations ranged 4-5 cm/s2, |r|=0.72), S1.2-1.6 (proportion of leaf speed ranged 1.2-1.6 cm/s, |r|=0.71), maximum leaf acceleration (LAmax, |r|=0.70) and leaf sequence variability (LSV, |r|=0.68) exhibited stronger GPRs correlations for ArcMap. Conversely, indexes related to beam area, including cross axis score (CAS, |r|=0.64), aperture area variability (AAV, |r|=0.57), small aperture score 5 mm (SAS5, |r|=0.56) and converted aperture metric (CAM, |r|=0.56) demonstrated stronger correlations for ArcCHECK.
Conclusions:
The impact of radiotherapy plan complexity on measurement verification displayed detector dependency, with detector size, spatial resolution, and geometric configuration critically influencing GPR-complexity correlations.