Optical and simulated visual outcomes of two presbyopia-correcting intraocular lenses tested in clinically relevant decentration and tilt conditions
Related Experiment Video
Updated: Jan 8, 2026

Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
Fátima Cuellar1, Fidel Vega1, David Madrid2
1Applied Optics and Image Processing Group, Optics and Optometry Department, Universitat Politècnica de Catalunya-BarcelonaTech, Violinista Vellsolà 37, 08222 Terrassa, Barcelona, Spain.
Abstract:
We evaluated the effect of intraocular lens (IOL) misalignment on image quality under clinically relevant conditions (0.2, 0.4 mm decentration and 2°, 4° tilt). Two presbyopia-correcting IOLs, an extended depth of focus (EDOF) (Alcon Vivity) and a diffractive multifocal (Alcon PanOptix), were tested. Through-focus modulation transfer function (MTF)-based measures were taken using an on-bench model eye with average corneal spherical aberration and apertures of 3.0 and 4.5 mm. Optical image quality and simulated visual acuity (VA) were assessed for both aligned and misaligned IOLs. The aligned EDOF lens showed superior distance vision quality compared to the multifocal, particularly with the 4.5 mm pupil. The EDOF lens was, however, more sensitive to decentration than the multifocal, which showed very low sensitivity across far, intermediate, and near regions with both apertures. Degradation with IOL tilt occurred at a slower rate than with decentration. Despite the optical degradation produced by mild misalignment (up to 0.4 mm decentration and 4° tilt), the predicted VA demonstrates that these effects should be negligible on postoperative vision with a 3 mm pupil. This holds true not only at a far distance, but also at the relevant intermediate and near foci.
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...