Related Experiment Video
Updated: Jan 8, 2026

Modified Blood Collection from Tail Veins of Non-anesthetized Mice with a Vacuum Blood Collection System and Eyeglass Magnifier
Published on: February 2, 2019
Evaluation of Blood-Sampling Performance of Automatic Fingertip-Blood-Sampling System and Correlation Between
Takashi Irie1, Kei Takenaka2,3, Sakuichiro Adachi1
1Healthcare Business Group, Diagnostic System Business, Hitachi High-Tech Corp., Hitachinaka, Ibaraki, Japan.
This study shows an automatic fingertip blood-sampling system (FBS-FV) precisely punctures skin and can predict blood volume using vessel imaging. Further development may enhance blood collection consistency.
Area of Science:
- Biomedical Engineering
- Medical Devices
- Diagnostic Technologies
Background:
- Accurate blood volume collection is crucial for diagnostics.
- Existing methods can be invasive or inconsistent.
- Novel automated systems are needed for improved blood sampling.
Purpose of the Study:
- Evaluate the performance of an automatic fingertip blood-sampling system with fingertip vessel-puncture function (FBS-FV).
- Examine the relationship between sampled blood volume and fingertip blood-vessel image features.
Main Methods:
- Conducted blood-sampling experiments with 18 participants.
- Utilized fingertip blood-vessel imaging for puncture site selection.
- Analyzed puncture accuracy, blood volume, and image features (V and C).
- Applied multiple regression to predict blood volume from image features.
Main Results:
- Achieved high puncture accuracy with deviation < 1 mm.
- Observed a 42% rate of exceeding the target blood volume (650 μL).
- Demonstrated predictive potential for blood volume using image features (R²=0.64 for high-volume, R²=0.41 for low-volume groups).
Conclusions:
- The FBS-FV offers precise puncture performance.
- Fingertip blood-vessel image features show potential for predicting blood volume.
- Optimization of compression control may enhance blood sampling consistency.