Related Experiment Video
Updated: Jan 8, 2026

Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
Predictive accuracy of CT-based planning in robotic-assisted total knee arthroplasty: A prospective observational
Waldo Gonzalez1,2, Jorge Isla1,3, Rafael Calvo1
1Faculty of Medicine, Clinica Alemana Santiago - Universidad del Desarrollo, Santiago de Chile, Chile.
Background:
The appropriate implant balance sizes are critical for successful total knee arthroplasty (TKA). Robot-assisted TKA (RA-TKA) utilizes preoperative CT-based planning to improve the precision of predicting implant sizes and necessary bone resections. We hypothesized that the RA-TKA system would accurately predict bone resections and component sizing while maintaining accurate alignment at the time of surgery.
Methods:
Prospective collection of data concerning the outcomes of RA-TKA surgeries conducted at our institution. Preoperative standard CT scans were employed to develop a three-dimensional (3D) model of the knee to estimate required bone resections for the femur and tibia and determine the most appropriate implant sizes. During the procedure, these images were aligned with the patient's anatomy to create a customized model, facilitating real-time evaluations of medial-to-lateral balance within the extension/flexion gap and the alignment of TKA components. We assessed the discrepancies between predicted and actual bone resections and implant sizes, measuring the postoperative mechanical axis of the lower limb. Statistical analyses were performed using STATA v.18.5, applying a 95 % confidence interval.
Results:
40 patients who underwent TKA due to severe osteoarthritis. The RA-TKA system accurately predicted the sizes of femoral and tibial components within one implant size in all cases (100 %). Out of the 240 evaluated bone cuts, 87.5 % deviated by less than 2 mm from the preoperative plan, indicating high accuracy. Significant differences were noted in the lateral posterior femoral condyle (p = 0.03) and lateral tibial plateau resections (p < 0.001), while other cuts showed minimal variation.
Conclusions:
The RA-TKA platform enhanced both preoperative planning and intraoperative performance while providing precise bone resections, correct component sizing, and favorable postoperative alignment.
Level Of Evidence:
Level IV.