Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Tacrolimus Assay Comparison: Analytical and Operational Considerations for Immunoassay vs Mass Spectrometry Testing
Rebekah Goldblatt1, Florencia Monge2, Cory Hansen1
1TriCore Reference Laboratories, Albuquerque, NM, United States.
The Roche Elecsys immunoassay offers a faster, simpler method for tacrolimus monitoring in transplant patients, showing excellent correlation with LC-MS/MS and significantly improving workflow efficiency.
Area of Science:
- Clinical Chemistry
- Transplant Medicine
- Analytical Chemistry
Background:
- Tacrolimus monitoring is crucial for transplant recipients.
- Liquid chromatography-tandem mass spectrometry (LC-MS/MS) offers accuracy but is labor-intensive.
- Immunoassays (IA) are faster but historically lacked specificity due to metabolite cross-reactivity.
Purpose of the Study:
- To compare the performance of the Roche Elecsys tacrolimus immunoassay with an in-house LC-MS/MS assay.
- To evaluate workflow improvements and turnaround times (TAT) after implementing the immunoassay.
Main Methods:
- A comparative analysis of 73 residual patient samples using LC-MS/MS and the Roche Elecsys immunoassay on a cobas® e801 analyzer.
- Statistical methods included Deming regression and Bland-Altman analysis.
- Evaluation of linearity, imprecision, workflow, and TAT.
Main Results:
- Excellent correlation was observed between the Roche Elecsys assay and LC-MS/MS (Deming regression: y = 1.004x + 0.4848).
- Bland-Altman analysis showed a bias of 0.52 ng/mL (8.1%).
- A 42%-51% improvement in TAT was achieved with the immunoassay.
Conclusions:
- The Roche Elecsys tacrolimus assay demonstrates excellent correlation, precision, and linearity compared to LC-MS/MS.
- Switching to the immunoassay significantly enhances workflow and reduces TAT in tacrolimus therapeutic drug monitoring.