Related Experiment Video
Updated: Jan 13, 2026

E-Patient Counseling Trial E-PACO: Computer Based Education versus Nurse Counseling for Patients to Prepare for Colonoscopy
Published on: August 1, 2019
Next-Generation SCS Programming Platform: Enhancing ECAP Fidelity and Objectivity to Improve Patient Experience
Daniel J Parker1, Ajay B Antony2, Gregory L Smith3
1Saluda Medical, Level 2/5 Eden Park Dr, Macquarie Park, NSW, 2113, Australia. daniel.parker@saludamedical.com.
Introduction:
Evoked compound action potentials (ECAPs) are neurophysiological biomarkers of neural activation during spinal cord stimulation (SCS). Clear distinction between ECAPs and nonphysiological signals is critical to the application of contemporary, ECAP-based closed-loop (CL) SCS therapies. Herein, we evaluated the performance and user acceptability of a novel programming software that automates generation of ECAP-based CL-SCS programs-the Assisted Programming Module (APM).
Methods:
We report results from two prospective, multicenter, single-arm, feasibility studies: Freshwater (NCT04662905) and Rosella (NCT06057480). APM performance was compared with the previous generation programming software and other published methods. Performance was assessed by comparing signal-to-noise ratios, artifact rejection, and other objective parameters. User acceptability was assessed using questionnaires administered to SCS users.
Results:
The APM successfully generated a CL program in 96% of initial programming sessions (n = 81/84; Freshwater, 31/34; Rosella, 50/50). In the Rosella study, median time to generate an automated CL program (n = 68) was 11.9 min [interquartile range (IQR) 9.9-14.0]. Median dose ratio was 1.31 (IQR 1.20-1.46) at end of trial (n = 24), 1.34 (IQR 1.13-1.51) at 1 month post implant (n = 16), and 1.32 (IQR 1.21-1.48) at 3 months post implant (n = 15). At least 90% of patients [trial, 90% (27/30); implant, 94% (17/18)] were satisfied with their programming experience, and ≥ 90% of patients [trial, 90% (26/29); implant, 94% (16/17)] felt in control of their therapy. The APM achieved a mean signal-to-noise ratio of 4.6 ± 1.2, a 35% improvement over the previous generation ECAP dose-controlled CL-SCS system. Detectable artifact leakage rates decreased by 75% when compared with other published methods without compromise to signal-to-noise performance.
Conclusions:
Next-generation ECAP dose-controlled CL technology demonstrated strong feasibility, high patient satisfaction and therapy control, and superior ECAP signal fidelity compared with existing methods. By standardizing CL-SCS programming and enhancing signal fidelity, the APM may improve workflow efficiency and long-term therapy outcomes in chronic pain management.
Trial Registration:
ClinicalTrials.gov identifiers: NCT04662905, NCT06057480.
More Related Videos
Related Concept Videos
Patient-centered Care
Nursing Clinical Information System
A Nursing Clinical Information System (NCIS) is a specialized type of healthcare information system tailored to meet the unique needs of nursing practice. It incorporates the principles of nursing informatics to streamline information management and improve the quality of care delivery.
Critical attributes of NCIS include:
Integrated Healthcare System
Nursing Implementation
The five steps to implementing effective nursing care include reassessing the patient, reviewing and revising the existing nursing care plan, organizing the resources and care delivery, anticipating and preventing complications, and implementing nursing interventions.
Standards of Care II
Methods of Documentation VI: Case Management Model
For example, a patient with a chronic...