Spatiotemporal Dynamics of Cochlear Implant Electrode Impedance in Standard and Dexamethasone-Eluting Electrode
Uzair A Khan1, Amy S Bussa1, Jacob Oleson2
1Department of Otolaryngology-Head and Neck Surgery, University of Iowa Health Care, Iowa City, IA.
Objective:
To evaluate the spatiotemporal dynamics of cochlear implant electrode impedance in standard and dexamethasone-eluting electrode arrays (Dex-EA).
Study Design:
Prospective, double-arm study.
Setting:
Single tertiary referral center.
Patients:
Thirty-one (17 control and 14 experimental) subjects meeting Food and Drug Administration criteria for cochlear implantation.
Interventions:
Standard-of-care cochlear implant with intraoperative postinsertion impedance measurement.
Main Outcome Measures:
Daily impedance measurements for a targeted 90 days postinitial activation.
Results:
This study demonstrates the spatiotemporal dynamics of electrode impedance observed over the first 90 days postimplantation with standard EAs, and the effects of Dex-EAs on these dynamics. For the control group, impedances for the 12 basal electrodes (1 to 12) increase over time while they decrease for the 10 apical electrodes (13 to 22). In comparison, not only do Dex-EAs result in smaller impedance values across all 22 electrodes over the first 90 days after initial activation, but the impedances also tend to decrease over time at all electrode locations. There is a statistically significant difference in impedance between the 2 cohorts at every electrode (P<0.0001), but the difference is largest for the basal electrodes. Dex-EAs also result in lower within-subject variability across repeated impedance measures over this short time window.
Conclusions:
Dex-EAs result in smaller, more stable impedance values across repeated measures over the first 90 days after initial activation as compared with commercially available standard CI electrode arrays. The dramatically lower impedances, reduced fluctuations, and resistance to rising values particularly at basal locations reflect the clinical advantages of Dex-EA toward diminishing the foreign body response and may also enhance stimulation from the CI.
Related Concept Videos
Susceptibility, Permittivity and Dielectric Constant
Potential Due to a Magnetized Object
The vector...
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
Ampere-Maxwell's Law: Problem-Solving
To solve the problem, we can use the equations from the analysis of an RC circuit and Maxwell's version of Ampère's law.
For the first part of the...
The Debye–Hückel Theory of Electrolyte Solutions
The Electrical Double Layer