Thermal intervention improves pulse oximetry accuracy in critically ill patients with low perfusion: a
Natalie Karlsson1, Felicia Olsson Lindstrand2, Lotta Johansson2,3
1Department of Thoracic Anaesthesiology and Intensive Care, Region Västra Götaland, Sahlgrenska University Hospital, Gothenburg, Sweden.
Background:
Pulse oximetry is essential for continuous oxygen monitoring in intensive care, yet its accuracy declines in patients with low peripheral perfusion, risking both unrecognised hypoxaemia and inappropriate oxygen therapy. The perfusion index (PFI) reflects peripheral blood flow and is often reduced in critically ill patients with impaired microcirculation. Simple bedside strategies to restore PFI and improve SpO₂ accuracy remain underexplored.
Methods:
In this prospective quasi-experimental study, 46 adult ICU patients with arterial catheters and baseline PFI < 1.0 underwent localised peripheral warming using a wrist-forearm heating pad for 15 min. The warming pad maintained a surface temperature of 54 °C, with directly measured skin-interface temperatures of 41-42 °C. SpO₂, PFI, and arterial oxygen saturation (SaO₂) were measured immediately before and after the intervention. The primary outcome was the change in PFI; the secondary outcome was the improvement in SpO₂ accuracy (SpO₂-SaO₂ bias).
Results:
Thermal intervention increased PFI from a median (IQR) of 0.56 (0.34-0.78) to 3.59 (2.45-4.77) (p < 0.001; Hedges' g = 2.43). The pre-intervention SpO₂-SaO₂ bias was 4.09% (95% limits of agreement 0.61-7.56%), which decreased to 0.00% (- 1.22-1.23%) after warming. Improvements were consistent across subgroups and unrelated to cardiac index, vasoactive use, or skin pigmentation.
Conclusions:
A brief, localised thermal intervention markedly improves peripheral perfusion and restores pulse-oximetry accuracy to within the clinically acceptable ± 2% range in critically ill patients with low PFI. However, ceiling effects at SpO₂ values near 100% and the pre-post design limit the strength of causal inference. This simple, non-invasive technique can be readily integrated into ICU practice to enhance the reliability of oxygen monitoring and reduce the risk of undetected hypoxaemia or hyperoxaemia.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...