Related Experiment Video
Updated: Jan 13, 2026

Comprehensive & Cost Effective Laboratory Monitoring of HIV/AIDS: an African Role Model
Published on: October 31, 2010
Comparative Performance of Flow Cytometry and ELISA for Detecting Ki-67 and HLA-DR in Whole Blood: Implications for
Jitendra S Dangi1, Mohita Pandey2, Anurag K Jain3
1Department of Surgery, Govt. Bundelkhand Medical College, Sagar, Madhya Pradesh, India.
Background:
This study aimed to compare the performance of flow cytometry and enzyme-linked immunosorbent assay (ELISA) in detecting Ki-67 and HLA-DR in whole blood from individuals with active TB, latent TB infection (LTBI), and healthy controls.
Methods:
A total of 60 participants were enrolled and categorized into three groups: active TB (n = 20), LTBI (n = 20), and healthy controls (n = 20). Whole-blood samples were analyzed using flow cytometry for intracellular Ki-67 and surface HLA-DR expression in T cells, and ELISA for quantification of their soluble forms. Statistical comparisons were performed using ANOVA and Pearson correlation.
Results:
Flow cytometry revealed significantly elevated Ki-67 (12.4 ± 2.1%) and HLA-DR (18.7 ± 3.6%) expression in active TB compared to LTBI (5.6 ± 1.4%, 9.5 ± 2.1%) and controls (1.9 ± 0.7%, 3.2 ± 1.2%), respectively (P < 0.001). ELISA also detected increased concentrations of both markers in active TB (Ki-67: 4.2 ± 1.1 ng/mL; HLA-DR: 6.8 ± 1.7 ng/mL) but with reduced sensitivity. A strong positive correlation was observed between flow cytometry and ELISA results for both biomarkers (r > 0.75, P < 0.01).
Conclusion:
Flow cytometry offers high sensitivity for TB immune markers, while ELISA is a practical alternative. Ki-67 and HLA-DR are promising biomarkers for TB diagnosis and monitoring.