Related Experiment Video
Updated: Jan 15, 2026

A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
Workflow, Alarm Notification and Technology Acceptance of Continuous, Wearable-Based Vital Sign Monitoring on A
Joan Alsolivany1,2, Lina Mosch3,4, Lars Wessels1,2
1Department of Neurosurgery.
Background:
Continuous, wearable-based vital sign monitoring can enhance patient safety and reduce intensive care demands but is not routinely used on standard neurosurgical wards. This study assessed alarm events and technology acceptance during the implementation of such a system for postoperative patients in a university hospital's neurosurgical ward.
Methods:
In this pilot feasibility study, selected patients were continuously monitored for 12 to 24 hours after elective neurosurgery on a 44-bed ward. Phase 1 (June to December 2019) used 12 mobile units with manual alarm documentation; Phase 2 (August 2023 to January 2024) used 44 fixed bedside units with automatic documentation. Alarm patterns and nurse-rated technology acceptance (4-point Likert scale) were analyzed.
Results:
Vital signs were monitored in 214 patients (median age: 58, IQR: 42 to 70) in Phase 1 and 290 patients (median age: 60, IQR: 43 to 77) in Phase 2. Procedures included cranial, spinal, and peripheral nerve surgeries. Phase 1 recorded 30 nonclinical alarms and no clinical alarms. In contrast, Phase 2 registered 14,500 clinical and 185,744 nonclinical alarms, mostly from detached sensors (177,989/185,744; 96%). Clinical alarms were mainly due to hypoxia (8305/14,500; 57%) and tachycardia (3487/14,500; 30%). Most alarms were acknowledged within 30 seconds (clinical: 12,969/14,500; 90% and nonclinical: 181,447/185,744; 98%), with delayed responses mostly for nonclinical events (3625/4189; 87%). Nurses reported improved convenience, usability, and system connection in Phase 2.
Conclusions:
Wearable-based vital sign monitoring is feasible on neurosurgical wards, but high nonclinical alarm rates highlight the need to refine alarm management strategies for effective clinical integration.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring