Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
Noise-optimized cone-beam computed tomographic imaging of different post materials with multiple settings: A
Solaleh Shahmirzadi1,2, Dilek Helvacioglu-Yigit3, Umut Seki4
1Division of Oral and Maxillofacial Radiology, Department of Diagnostic Sciences, Texas A&M School of Dentistry, Dallas, USA.
Optimizing cone-beam computed tomography (CBCT) settings, including metal artifact reduction (MAR) and kilovoltage peak (kVp), enhances image quality for visualizing dental posts. Higher contrast-to-noise ratio (CNR) is achieved with specific settings, improving diagnostic accuracy.
Area of Science:
- Dental Imaging
- Radiology
- Biomaterials
Background:
- Accurate visualization of endodontic posts is crucial for diagnosis.
- Metal artifacts in cone-beam computed tomography (CBCT) can obscure details.
- Optimizing imaging parameters is necessary to improve image quality.
Purpose of the Study:
- To evaluate the impact of different post materials on contrast-to-noise ratio (CNR).
- To assess the influence of various CBCT imaging settings and metal artifact reduction (MAR) algorithms on CNR.
- To determine optimal imaging parameters for visualizing teeth with posts.
Main Methods:
- Forty single-rooted teeth with four different post materials were scanned using a Planmeca ProMax 3D Max.
- Scans were performed at varying kilovoltage peak (kVp), MAR levels, and voxel sizes.
- Image analysis using ImageJ software calculated CNR; effective dose was also determined.
Main Results:
- Higher CNR was observed in the middle root third at 96 kVp, 200 μm voxel size, and medium/high MAR.
- Gold posts exhibited the lowest mean CNR; other materials showed no significant differences.
- All tested factors significantly influenced mean CNR (P<0.05).
Conclusions:
- A combination of low resolution (200 μm voxel size), medium MAR, and 96 kVp yielded the highest CNR and image quality.
- These settings improved visualization of teeth with post materials using the Planmeca ProMax.
- This optimized protocol also resulted in one of the lowest effective doses.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...