Detection of PD-1 and CD28 Expression in Lymphocytes by Flow Cytometry
Shuying Chen1,2, Xiang Wu1,2, Yiting Tang1
1Department of Laboratory Medicine, Huashan Hospital, Fudan University, Shanghai, China.
The Journal of Applied Laboratory Medicine
|January 16, 2026
Summary
A new flow cytometry method accurately detects immune checkpoints PD-1 and CD28 on lymphocytes. This reliable tool supports cancer immunotherapy research and clinical decision-making.
Area of Science:
- Immunology
- Biotechnology
- Clinical Diagnostics
Background:
- Immune checkpoint analysis is crucial for cancer immunotherapy research, immune microenvironment exploration, and biomarker discovery.
- Programmed Death 1 (PD-1) and CD28 are key T cell surface receptors regulating immune responses.
- Accurate detection of immune checkpoints like PD-1 and CD28 on lymphocytes is vital for understanding immune responses and guiding cancer immunotherapy.
Purpose of the Study:
- To develop and optimize a flow cytometry-based method for quantitative assessment of PD-1 and CD28 expression on lymphocytes.
- To establish and validate a detailed protocol for immune checkpoint detection using the CYTEK NL-CLC flow cytometer.
Main Methods:
- Utilized the CYTEK NL-CLC flow cytometer for quantitative detection of PD-1 and CD28 on lymphocytes.
- Developed and validated a flow cytometry protocol, assessing key performance parameters.
Main Results:
- The developed method demonstrated high sensitivity and specificity for immune checkpoint detection.
- Validation metrics including precision, linearity, stability, reference interval, and accuracy met acceptable criteria.
- Results are approved for clinical testing, indicating the method's readiness for application.
Conclusions:
- The CYTEK NL-CLC flow cytometer provides a reliable platform for immune checkpoint analysis.
- This method supports cancer immunotherapy workflows and personalized medicine strategies.
- The validated assay is suitable for both clinical and research settings.