A-Maze-Ox: a novel gas-exchange-area-adjustable oxygenator for extremely preterm infants-design and proof of concept
Franziska Schubert1, Jan Heyer2, Maximilian Lunemann2
1Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany. schubert@ame.rwth-aachen.de.
Background:
Each year, over half a million children worldwide are born extremely prematurely (EPI), often requiring potentially harmful ventilatory support. An artificial placenta with gas exchange provided by an oxygenator offers an alternative. However, current oxygenators are not designed for long-term applications in patients experiencing growth. Thus, a new type of "growing" oxygenator is needed.
Methods:
We developed the A-Maze-Ox, a novel maze-inspired membrane oxygenator featuring two concentric compartments that can be opened sequentially to address patient growth. Each compartment has a priming volume of 5 mL and a gas exchange area of 0.065 m². The prototype was tested for feasibility, gas transfer performance and pressure loss according to ISO 7199.
Results:
Adding the second compartment increased O2 and CO2 transfer performance by 57.06% and 35.59%, respectively. While CO2 transfer was mostly sufficient, O2 transfer remained below the target. The targeted maximum pressure drop of 20 mmHg was exceeded at 90 mL/min.
Conclusions:
The A-Maze-Ox demonstrates the ability to adapt to increasing demands with good results in CO2 elimination but requires improvement in terms of O2 transfer and pressure drop. However, it shows the potential for a growing oxygenator in future artificial placenta applications.
Impact:
Current oxygenators cannot adapt to the dynamic patient growth of extremely premature infants. A-Maze-Ox-a novel membrane oxygenator with two compartments can adapt to growth. Each compartment has a volume of 5 mL and a gas exchange area of 0.065 m². Oxygen transfer increased by 57.06% and carbon dioxide transfer by 35.59% when adding the second compartment. A-Maze-Ox could improve health of extremely premature infants in the future.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...