Can new ultrasound imaging techniques improve breast lesions characterization?: prospective comparison between
Giacomo Bonito1, Maurizio Renda2, Silvia Gigli3
1Department of Radiological, Oncological and Pathological Sciences, Policlinico Umberto I, Sapienza University of Rome, Viale Regina Elena 324, 00161, Rome, Italy. giacomo.bonito@uniroma1.it.
Background:
Multiparametric ultrasound (mpUS) integrates BI-RADS (Breast Imaging Report And Data System) morphological assessment with biomechanical biomarkers, including strain elastography (SE) and shear-wave elastography (SWE), along with microvascular flow (M-Flow). Furthermore real-time computer-aided diagnosis (CAD/AI) systems, including Live Breast Assist and S-Detect are employed to standardize image interpretation, thereby reducing operator dependence.
Purpose:
To evaluate the diagnostic accuracy and reproducibility of mpUS supported by CAD/AI versus B-mode ultrasound alone, assessing their concordance with histopathology for the benign/malignant dichotomy and the incremental contribution of SE, SWE, and M-Flow. Additionally inter-operator agreement was quantified.
Materials And Methods:
In this single-center prospective study 78 consecutive patients presenting with 78 breast lesions, underwent B-mode/BI-RADS evaluation, Color Doppler, M-Flow (percentage of vascularization), SE (strain ratio), and SWE (quantitative stiffness in kPa). Live Breast Assist was semi-automatically employed during data acquisition, while S-Detect provided focal lesion characterization when deemed necessary. Histopathological examination served as the gold standard.
Results:
Histology identified 60 malignancies (76.9%) and 18 benign lesions (23.1%).The highest diagnostic performance was shown by M-Flow (AUC 0.926), reaching 95.1% of sensitivity and 82.4% of specificity. SE also showed strong discriminative ability (AUC 0.885). SWE exhibited lower accuracy (AUC 0.658). B-mode alone achieved an AUC of 0.864. Although with a higher false-negative rate compared with the expert operator, Live Breast Assist demonstrated high specificity (88.2%) and an excellent PPV (96.4%). Inter-operator agreement was substantial (κ = 0.63).
Conclusion:
mpUS with CAD/AI support enhanced diagnostic accuracy and reproducibility for benign/malignant distinction, providing robust biomechanical and vascular biomarkers. This approach may reduce unnecessary biopsies in indeterminate BI-RADS lesions, though tumor grading remains the purview of pathology.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...