Atrioventricular synchrony maintained by a dual-chamber leadless pacemaker in real-world conditions
Pascal Defaye1, Vivek Y Reddy2, James E Ip3
1Department of Cardiology, CHU Grenoble Alpes, Grenoble, France.
Aims:
A dual-chamber leadless pacemaker (LP) system relies on a beat-to-beat, wireless, implant-to-implant (i2i™) communication between implanted devices in the right atrium and right ventricle (A-to-V and V-to-A). Atrioventricular (AV) synchrony was evaluated during an ambulatory 24 h period of daily living to determine if its effectiveness can be maintained under real-world conditions.
Methods And Results:
A prospective, single-arm, multicentre clinical study (ClinicalTrials.gov identifier NCT05252702) evaluated the overall safety and performance of the dual-chamber LPs in patients with standard indications for dual-chamber pacing. Following the 3-month visit, eligible patients wore 12-lead Holter monitors for 24 h while resuming the activities of daily living while under clinically driven programmed settings. Leadless pacemaker diagnostic data were interrogated when patients returned to the clinic. An independent Holter core laboratory adjudicated the proportion of synchronous beats (PR intervals within paced and sensed AV delay bounds). Relationships between AV synchrony and i2i communication success rate, implant indication, heart rate, and AV event type were evaluated. Among 47 analysable patients, mean AV synchrony was achieved in 96.7% of beats, exceeding mean A-to-V and V-to-A transmission success rates of 93.4% and 92.1%, respectively. There were no significant differences in AV synchrony between sinus node and AV block patients. Across all AV paced/sensed cycle combinations and heart rate ranges (including >100 b.p.m.), AV synchrony was sustained >90% in an ambulatory real-world setting.
Conclusion:
A dual-chamber leadless pacing system demonstrated >96% overall AV synchrony for a 24 h period of daily living while programmed to clinically appropriate settings, thus showing that true dual-chamber DDD(R) pacing can be maintained in a real-world environment.
Related Concept Videos
Real-World Application of Classical Conditioning
Higher-order, or second-order, conditioning occurs when a neutral stimulus becomes associated with an already established conditioned stimulus through repeated pairings. For instance, if a dog has been...
Design Example: Maintaining Level of an Embankment
Chambers of the Heart
Deoxygenated blood from the body is received in the right...
Role of Vitamins in Maintaining Bone Health
Vitamin A
Vitamin A is involved in the process of bone remodeling. Retinoic acid, the active metabolite of Vitamin A, has nuclear receptors in osteoblasts and osteoclasts, which are involved in bone remodeling.
Vitamin B12
Vitamin B12 acts as a cofactor during the formation of osteoblast-related proteins, such as osteocalcin. Vitamin B12 plays a role...
Conditions on Early Earth
Classical Conditioning
Ivan Pavlov observed that dogs...