Synchrotron split-beam method for clinical photon-counting computed tomography
Stevan Vrbaški1, Sandro Donato2, Elena Longo3
1Elettra-Sincrotrone Trieste S.C.p.A, 34149 Basovizza Trieste, Italy; Department of Physics, University of Trieste, Via Valerio 2, 34127 Trieste, Italy; Carl E. Ravin Advanced Imaging Laboratories and Center for Virtual Imaging Trials, Duke University, Durham, NC 27705, USA; University of Novi Sad, Faculty of Medicine, Hajduk Veljkova 1-3, 21000 Novi Sad, Serbia.
Purpose:
Free-space phase contrast propagation coupled with a photon-counting detector enables CT imaging with improved contrast in soft tissues at lower radiation dose. In addition, photon-counting detectors enable inherent spectral separation that can be used to capture tissue contrast at different energy levels. The objective of this study was to (i) develop a novel split-beam method for spectral synchrotron-based imaging considering limitations of photon-counting technology and clinical requirements, and (ii) propose a redefined mathematical model to calculate contrast-to-noise ratio in spectral imaging applications.
Methods:
Our novel approach was applied in a CT imaging setup using a custom-made breast phantom with tissue-equivalent inserts and compared to the more common setup utilizing monochromatic beams. To complement the traditional contrast-to-noise ratio metric, a new mathematical framework for spectral contrast-to-noise ratio was introduced as a composite metric that integrates the signal-to-noise performance across spectral channels.
Results:
The results show that the split-beam method proposed in this study obtains a comparable spectral contrast-to-noise ratio at the same radiation dose. Relative spectral contrast-to-noise differences were 0.12 (polyethylene), 1.92 (polyamide), 1.19 (polymethylmethacrylate), -0.29 (polyoxymethylene), and -0.17 (polytetrafluoroethylene) when comparing spectral imaging with two monochromatic beams at energies of 24 keV and 38 keV against the split-beam method.
Conclusion:
The potential advantages of the split-beam method for spectral CT imaging are numerous - it avoids non-rigid deformations, is fast to implement, and enables optimization in the post-processing step. The model for contrast-to-noise ratio redefined in this study applies to new generation spectral CT scanners beyond synchrotron setups.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Area Computation by the Alternative Coordinate Method
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
¹H NMR: Complex Splitting
Splitting diagrams or splitting tree diagrams are routinely used to depict such complex couplings. While drawing splitting diagrams, the splitting with the larger coupling constant is usually applied...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...