Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
The dual gap ionization chamber: a novel ionization chamber design for reference dosimetry to automatically correct
Marina Orts1, Séverine Rossomme2, Kevin Souris2
1UCLouvain, Institut de Recherche Expérimentale et Clinique, Molecular Imaging Radiotherapy and Oncology (MIRO), Brussels, Belgium.
Abstract:
Objective.Current dosimetry protocols typically recommend multiple measurements to determine recombination correction factors (ks), increasing the time required for dose measurements in the quality assurance workflow. We propose a novel dual-gap ionization chamber (DGIC) design for reference dosimetry featuring two air gaps of different thicknesses within a single device. This design enables the determination ofksdirectly from the same measurements required to determine absorbed dose-to-water. Thus, eliminating the need for separate measurements to correct for recombination losses. The approach relies on analyzing the charge ratio between the two gaps, under ultra high dose rates (UHDR) and dose per pulse (DPP) under ultra high DPP (UHDPP) conditions.Approach.A DGIC prototype with electrode distances of 1 and 0.6 mm was developed and tested using different beam qualities: (1) a 240 MeV u-1clinical carbon ion beam at conventional field dose rates of 25 Gy min-1, (2) a 226 MeV continuous proton beam with a current between 5 and 800 nA at the cyclotron exit, where the maximum approximately corresponds to 200 Gy s-1in the treatment room and (3) a 9 MeV electron beam with a DPP from 0.03 to 4.2 Gy, a frequency of 60 Hz and a pulse duration between 0.7-3.9μs.ks-factors were derived for the top cavity of 1 mm gap using the DGIC method and compared against the following: for proton and carbon ions, comparisons were made with the Jaffé plot method. For the electron beam, it was compared with a dose rate independent device, a flashDiamond detector, and the integrated current transformer of the LINAC.Main results.A DGIC prototype was able to successfully correct for recombination losses under different beam modalities: for initial recombination in a clinical carbon ion beam, volume recombination in UHDR proton beam with field dose rates of 200 Gy s-1and in UHDPP electron beams, where four pulses were delivered with DPP up to 4.2 Gy (this DPP corresponds to an effective pulse duration of 3.9μs).Significance.A DGIC design and its inherent method provides a practical and accurate way of determining dose and dose rate in emerging radiotherapy treatment modalities.
Related Concept Videos
Ionization Energy
Chambers of the Heart
Deoxygenated blood from the body is received in the right...
Chemical Ionization (CI) Mass Spectrometry
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Matrix-Assisted Laser Desorption Ionization (MALDI)
Automatic Processing and Automatic Social Behavior