Effect of Monoenergetic Reconstructions and Ultra-High-Resolution on the Detection of Small Endoleaks With
Background And Objectives:
Following endovascular repair of an aortic aneurysm, patients need regular surveillance using CT to exclude endoleaks. The optimal technical parameters for photon-counting detector CT (PCD-CT) have not yet been established in this clinical context. This study examines the impact of the ultra-high resolution (UHR) mode and monoenergetic reconstructions at various keV on the detection of small endoleaks in experimental settings.
Materials And Methods:
An aneurysm phantom model mimicking medium- and large-sized patients was scanned using PCD-CT. Monoenergetic images (MEIs) were reconstructed with standard resolution at 40, 45, 50, 55, and 60 keV and with UHR at 45, 50, and 55 keV. Three independent blinded radiologists assessed 80 CT series containing 2880 simulated endoleaks with diameters of 2, 4, and 6 mm. The observers' ratings were analyzed and compared with the jackknife alternative free-response operational characteristics (JAFROC1) method by calculating a figure-of-merit for the performance ( A1 ). The Spearman R was calculated for the correlation between A1 and objective image quality.
Results:
The best readers' performance was achieved with the 50 keV UHR, 55 keV standard, 55 keV UHR, and 60 keV standard series ( A1 =0.66 to 0.72) in the medium and with the 55 keV standard, 55 keV UHR, and 60 keV standard reconstructions ( A1 = 0.40 to 0.49) in the large phantom. UHR provided higher A1 than the standard reconstruction at 50 keV for 4 to 6 mm endoleaks ( A1 , 0.82 vs. 0.72; P =0.015 in medium and 0.42 vs. 0.30; P =0.028 in large phantom) and for 6 mm endoleaks ( A1 , 0.93 vs. 0.83; P =0.002 in medium and 0.53 vs. 0.36; P =0.013 in large phantom). In both phantoms, readers' performance correlated negatively with image noise (Spearman R =-0.83; P =0.01 in the medium and Spearman R =-0.98; P <0.001 in the large phantom) and showed no correlation with the contrast-to-noise ratio of endoleaks ( P =0.91 and 0.73).
Conclusions:
The 55 keV standard or UHR and 60 keV standard reconstructions yielded the best performance for detecting small endoleaks with PCD-CT. The UHR mode improved performance at 50 keV for 4 to 6 mm and 6 mm endoleaks.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
14:09High-Throughput Total Internal Reflection Fluorescence and Direct Stochastic Optical Reconstruction Microscopy Using a Photonic Chip
Published on: November 16, 2019
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...