Sufficient stability using retrograde cannulated screws in a metacarpal fracture model: a biomechanical evaluation
Maximilian Heilig1, Julian Wagenhäuser2, Henner Huflage3
1Department of Trauma and Orthopaedic Surgery, BG Unfallklinik Frankfurt am Main, Frankfurt am Main, Germany.
Introduction:
The treatment of metacarpal shaft fractures using cannulated compression screws has become a viable method of osteosynthesis in recent years. However, most biomechanical studies focus primarily on the ultimate failure load, comparing it to plating or K-wires. There is limited biomechanical data how cannulated compression screws perform under realistic cyclic loads encountered during postoperative physiotherapy in comparison to plating.
Methods:
Oblique shaft fractures were created in both porcine metatarsal and human metacarpal bones. Fractures were reduced and treated with either a partially or fully threaded cannulated compression screw or a 2.0 TriLock hand plate. Cyclic loading was then applied for 3,000 cycles, ranging from 10 to 80 N, followed by an ultimate load-to-failure test.
Results:
The displacement after cyclic loading was lower for specimens being treated with a 2.0 TriLock hand plate compared to those treated with an intramedullary compression screw. In the subsequent load-to-failure tests, specimens with a fully threaded intramedullary screw withstood higher forces than those with a partially threaded screw or a plate. Notably, no superiority of the plate was observed in terms of ultimate loading in both porcine and human specimens.
Discussion:
The results indicate that using a cannulated compression screw for the treatment of metacarpal oblique fractures provides good biomechanical stability under cyclic loading. Additionally, the use of a fully threaded screw appears to increase the ultimate failure load. This study adds to the biomechanical evidence supporting the use of cannulated compression screws in the treatment of metacarpal oblique fractures. Surgeons should prioritize achieving the best and longest cortical anchorage when selecting the implant to optimize the patient's outcome.
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Nuclear Stability
To hold positively charged protons together...
RNA Stability
Stability
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...