Successful Extraction of Entrapped CIED Lead After TTVR
Johannes Kirchner1, Barbara Rubinic1, Muhammed Gerçek1
1Department of General and Interventional Cardiology/Angiology, Herz- und Diabeteszentrum NRW, Ruhr-Universität Bochum, Bad Oeynhausen, Germany.
Background:
Transcatheter tricuspid valve replacement (TTVR) has emerged as a therapeutic option for patients with symptomatic severe tricuspid regurgitation. Right ventricular (RV) leads from pre-existing cardiac implantable electronic devices are entrapped during TTVR device implantation.
Case Summary:
An 83-year-old woman with a dual-chamber cardiac implantable electronic device underwent TTVR in 2022. The procedure resulted in entrapment of the RV lead. After TTVR, the increase in lead impedance raised concern for lead failure, prompting consideration of a leadless pacemaker. During this procedure, the entrapped lead was extracted without causing malposition of the TTVR device.
Discussion:
Extraction of entrapped leads after TTVR may be feasible in experienced centers after cautious deliberation. Nevertheless, pre-emptive lead extraction should be considered to minimize the risk of unsuccessful extraction or serious complications.
Take-Home Message:
Extraction of entrapped RV leads without causing TTVR device malposition may be feasible, and leadless pacemakers can be considered a technically viable alternative for RV pacing in TTVR.
Related Concept Videos
Ecological Succession
Phase-lead and Phase-lag Controllers
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Extraction: Effects of pH
Extraction: Advanced Methods
Extraction: Partition and Distribution Coefficients
For extracting a solute from an aqueous phase into an...