Related Experiment Video
Updated: Jan 25, 2026

Whole-cell Currents Induced by Puff Application of GABA in Brain Slices
Published on: October 12, 2017
Increased brain coverage and efficiency when measuring current-induced magnetic fields by use of simultaneous
Teresa Cunha1,2, Fróði Gregersen1,2, Lars G Hanson1,2
1Section for Magnetic Resonance, DTU Health Tech, Technical University of Denmark, Kgs Lyngby, Denmark.
Purpose:
Magnetic resonance current density imaging (MRCDI) can non-invasively validate electric field simulations in volume conductor head models. Weak electric currents are injected using scalp electrodes while measuring the MR phase perturbations caused by the tiny magnetic fields (1-2 nT) induced by the current flow in tissue. MRCDI generally has a low signal-to-noise ratio, making it susceptible to technical imperfections and physiological noise. In this technical note, we tested and optimized simultaneous multi-slice (SMS) EPI for time-efficient and robust brain MRCDI.
Methods:
MRCDI data was acquired in a phantom and five human brains using SMS-EPI optimized for measuring current-induced phase perturbations. Multiband factors and interslice gaps were systematically varied and the resulting image quality assessed. In particular, the impact of interslice signal leakage on the measured phase was tested.
Results:
Current-free acquisitions showed the expected noise amplification with decreasing interslice distances. However, physiological noise generally dominated the human data, masking potential SMS-related penalties and making the overall noise levels identical to single-slice EPI for interslice gaps of at least 12 mm and multiband factors between 3 and 5. Upon application of electric currents, the phantom data revealed subtle artifacts for multiband factors 5 and 6, even for large gaps. Nevertheless, artifacts were absent in the human brain for multiband factors up to 5, where the performance of SMS-EPI approached that of single-slice measurements for sufficient interslice distances.
Conclusion:
Optimized SMS-EPI with multiband factors up to 5 and minimum interslice gaps of 12 mm performs on par with single-slice EPI, making it attractive for increasing brain coverage in MRCDI.
More Related Videos
Related Concept Videos
Magnetic Field Of A Current Loop
Force On A Current Loop In A Magnetic Field
Torque On A Current Loop In A Magnetic Field
Consider a rectangular current-carrying loop containing N turns of wire, placed in a uniform magnetic field. The net force on a current-carrying loop...
Crystal Field Theory - Tetrahedral and Square Planar Complexes
Crystal field theory (CFT) is applicable to molecules in geometries other than octahedral. In octahedral complexes, the lobes of the dx2−y2 and dz2 orbitals point directly at the ligands. For tetrahedral complexes, the d orbitals remain in place, but with only four ligands located between the axes. None of the orbitals points directly at the tetrahedral ligands. However, the dx2−y2 and dz2 orbitals (along the Cartesian axes) overlap with the ligands less than the dxy,...
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...

