Commissioning of a Monte Carlo-based scanning proton beam for breast cancer: Incorporating LETd calculations and
Zhen Cao1,2, Qing Zhang3, Jingfang Zhao1
1Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai Key Laboratory of Radiation Oncology, Shanghai Engineering Research Center of Proton and Heavy Ion Radiation Therapy, Shanghai, People's Republic of China.
Background:
Using a constant relative biological effectiveness (RBE = 1.1) in proton therapy may underestimate the RBE-weighted dose in high linear energy transfer (LET) regions at the distal end of the beam, thereby limiting the ability to accurately predict clinical outcomes.
Purpose:
To commission and validate a Monte Carlo (MC) model incorporating variable RBE for breast cancer proton therapy, enabling improved RBE-weighted dose calculation.
Methods:
A FLUKA-based MC model of a raster scanning proton beamline was commissioned and benchmarked against the clinically employed treatment planning system (TPS) (Siemens Syngo) and physical measurements. Dose-averaged LET (LETd) and variable RBE-weighted dose distributions were computed using McMahon (McM), McNamara (McN), and Wedenberg (Wed) models. Treatment plans for four representative breast cancer cases were recalculated to compare TPS and MC results using dose-volume histograms (DVH) and three-dimensional gamma (γ) analysis. LETd-volume histograms (LVH) and variable RBE-weighted dose distributions were analyzed to compare cases without adverse effects versus those presenting rib fractures or radiation pneumonitis.
Results:
The FLUKA-MC model showed good agreement with both the TPS results and the measured data, exhibiting proton range deviations within ±0.1 mm. The γ pass rates for the four patients are 94.0%, 92.2%, 92.6%, and 86.7%, respectively. LETd analysis of 0.5 cm3 volumes of rib revealed numerical differences (fracture cases: 11.1 and 10.8 keV/µm; non-fracture: 9.2 and 10.0 keV/µm). The RBE-weighted dose to 0.5 cm3 of the ribs was consistently elevated in fracture cases across all models (RBE = 1.1: 46.2-49.0 Gy; McM: 54.6-56.5 Gy; McN: 51.0-53.3 Gy; Wed: 50.6-52.5 Gy) versus non-fracture cases (RBE = 1.1: 44.0-45.3 Gy; McM: 52.2-53.8 Gy; McN: 48.6-50.1 Gy; Wed: 48.3-49.8 Gy). The estimated RBE values in the rib region were 1.60 (McM), 1.38 (McN), and 1.44 (Wed), which were derived from the mean LETd within 0.5 cm3 rib volumes. The RBE-weighted lung V20 was elevated in pneumonitis patients across all models. All variable RBE models predicted elevated RBE-weighted doses in distal proton beam regions across cases.
Conclusions:
The commissioned MC framework demonstrated the feasibility of integrating multiple variable RBE models for RBE-weighted dose estimation in proton therapy.
Related Concept Videos
Titration Calculations: Weak Acid - Strong Base
For the titration of 25.00 mL of 0.100 M CH3CO2H with 0.100 M NaOH, the reaction can be represented as:
Titration Calculations: Strong Acid - Strong Base
A titration is carried out for 25.00 mL of 0.100 M HCl (strong acid) with 0.100 M of a strong base NaOH. The pH at different volumes of added base solution can be calculated as follows:
(a) Titrant volume = 0 mL. The solution pH is due to the acid ionization of HCl. Because this is a strong acid, the ionization is complete and the hydronium ion molarity is 0.100 M. The pH of the solution is then:
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Calculating pH Changes in a Buffer Solution
Calculating the Equilibrium Constant
For example, gaseous nitrogen dioxide forms dinitrogen tetroxide according to this equation:
Leaky Scanning