Impedance Problems and Their Causes-A Single-Center Analysis of 601 Patients with De Novo Deep Brain Stimulation
Thomas Fortmann1,2,3, Samer Zawy Alsofy2,3, Antonio Santacroce2,3,4
1Department of Stereotactic Neurosurgery, St. Barbara Hospital Hamm, 59073 Hamm, Germany.
Abstract:
Background/Objectives: Patients with deep brain stimulation (DBS) require regular follow-up. When a sudden loss of therapeutic effect occurs, impedance abnormalities are often the underlying cause. If reprogramming cannot restore clinical benefit, revision surgery may be necessary to replace defective hardware. Since all three major manufacturers are used at our center, we analyzed our patient cohort to determine the incidence and causes of impedance abnormalities. Methods: All 601 patients who underwent de novo DBS implantation in Hamm between 2009 and 2025 were evaluated for impedance abnormalities. In cases requiring revision surgery, the specific cause was identified. The manufacturer, electrodes, and contacts involved were systematically analyzed. Results: A total of 25 of 601 patients required revision surgery. Revision rates were 2.67% in patients with Parkinson's disease, 6.19% in those with a tremor, and 5.71% in those with dystonia. Across manufacturers, 7.6% of patients with a Medtronic system required revision surgery, compared with 3.4% of patients with an Abbott system and no patients with a Boston Scientific system. The primary causes of revision were electrode-related problems (19/25), followed by extension defects (6/25), connector issues (4/25), and, in one case, a generator defect (1/25). Conclusions: Only 4.16% of patients required revision surgery due to impedance abnormalities. Patients with a tremor and non-segmented electrodes showed a higher incidence than those with Parkinson's disease or dystonia. Predominantly older Medtronic systems had the highest revision rate, whereas no Boston Scientific systems required revision. In most cases, the electrodes were the primary source of impedance abnormalities. A total of 52% of revisions were performed within two years and 92% were performed within six years of implantation.
Related Concept Videos
Patient-centered Care
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Impedance Combination
Center of Gravity
Center of Gravity
To determine its location, the principle of moments can be utilized by dividing the object into...
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...