Ultrasound-based Pulse Wave Velocity Evaluation in Mice
Published on: February 14, 2017
Reproducibility and agreement of pulse wave velocity and augmentation index over repeated assessments using two
Wesley Torres1,2, Earric Lee3,4, Emilia Laitinen2
1Laboratory of Investigation in Exercise (LIVE), São Paulo State University (UNESP), School of Technology and Sciences, Presidente Prudente, Brazil.
Objectives:
We investigated the agreement between pulse wave velocity (PWV) and augmentation index (AIx%) obtained in a controlled fasted condition versus a non-fasted, uncontrolled, "real-world" condition. Thereafter, we assessed the reproducibility of PWV and AIx% over three repeated visits under controlled fasted conditions.
Methods:
PWV and AIx% were assessed in one uncontrolled visit, and three controlled fasted visits after a 10-min supine rest in 28 adolescents (61% girls) aged 12-14 years.
Results:
Intraclass correlation coefficients (ICCs) between controlled visits were 0.5-0.6 (PWV) and -0.3-0.6 (AIx%) for PulsePen, and 0.5-0.6 (PWV) and 0.4-0.7 (AIx%) for Arteriograph. ICCs between a single uncontrolled visit and controlled visits were lower, especially for the Arteriograph (PWV: 0.0-0.1; AIx%: -0.4-0.0). Device agreement was poor (PWV ICC: -0.2-0.1; AIx% ICC: -0.5-0.3). Reproducibility using PulsePen was (PWV: 0.2-0.8; AIx%: -0.3-0.6) and Arteriograph (PWV: 0.0-0.1; AIx%: -0.4-0.0).
Conclusion:
The PulsePen had better reproducibility of PWV than the Arteriograph. Both devices had limited reliability and high variability in AIx% measurements. PWV from the PulsePen may be applicable in healthcare, while results from the Arteriograph should be interpreted with caution. Because of the poor agreement between the devices, the data from these devices should not be directly compared. These results may not be generalizable based on the modest sample size. In addition, the variability in real-world situations should not be overlooked.
Related Concept Videos
Velocity and Acceleration of a Wave
The velocity of the particles can be obtained by taking the partial derivative of the position equation with respect to time....
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Pulse Assessment Sites
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Pulse
The pulse serves as a clinical...