Timing and characteristics of carotid microembolic signals during pentaspline pulsed field ablation
Mitsuru Takami1, Kimitake Imamura2, Hidehiro Iwai1
1Division of Cardiovascular Medicine, Department of Internal Medicine.
Background:
Although pulsed field ablation (PFA) is generally considered safe, the mechanisms underlying both symptomatic and silent embolic events remain unclear.
Objectives:
This study aimed to investigate the timing, number, size, and characteristics of microembolic signals (MESs) during each step of the pentaspline PFA procedure using real-time carotid Doppler monitoring.
Methods:
MESs were assessed during ablation procedures in 32 patients, including 27 who underwent pulmonary vein (PV) isolation with the pentaspline PFA system and 5 control patients who received radiofrequency ablation limited to the right atrium.
Results:
The PFA group showed significantly more MESs than the control group. MESs were detected throughout the PFA procedure, with the highest counts observed during PFA delivery to the right PVs. Bubble signals were seen in all procedural steps. In contrast, high-intensity transient signals, which are mainly composed of solid components, were more common during PFA energy delivery. All 303 overcurrent errors-events in which energy delivery is stopped owing to overcurrent detection-occurred during basket-type PFA applications and were associated with a marked increase in MESs. Basket-configuration energy delivery resulted in a significantly higher number of MESs than the flower configuration, particularly in the right PVs. The size of MESs was significantly larger during PFA energy delivery than during catheter or sheath manipulation.
Conclusions:
This study demonstrated the characteristic timing, frequency, and morphology of MESs during pentaspline PFA procedures, providing insights that may help reduce embolic risks in future practice.
More Related Videos
Related Concept Videos
Basic Continuous Time Signals
The unit step function, denoted u(t), is zero for negative time values and one for positive time values, exhibiting a discontinuity at t=0. This function often represents abrupt changes, such as the step voltage introduced when turning a car's...
Basic Discrete Time Signals
The unit impulse or sample sequence is mathematically expressed as zero for all n values except at n=0, where it is one. The unit impulse sequence, denoted by δ(n), is the first difference of the unit step sequence, while the unit step sequence u(n) is the...
Sampling Continuous Time Signal
In the...
Characteristics of Life
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...