Megavoltage computed tomography image quality assessment in a modern tomotherapy system: a frequency-domain analysis
Related Experiment Video
Updated: Feb 4, 2026

Computed Tomography-guided Time-domain Diffuse Fluorescence Tomography in Small Animals for Localization of Cancer Biomarkers
Published on: July 17, 2012
Hiroshi Kunitomo1, Masato Yamada2, Yuta Eguchi3
1Medical Equipment Engineering, School of Medical Sciences, Fujita Health University, 1-98 Dengakugakubo, Kutsukake-cho, Toyoake, Aichi, 470-1192, Japan. hiroshi.kunitomo@fujita-hu.ac.jp.
Abstract:
This study aimed to comprehensively evaluate the image quality of iterative (IR) and filtered back-projection (FBP) reconstruction in a modern megavoltage computed tomography (MV-CT) system using frequency-domain methods in both the in-plane and longitudinal directions. MV-CT scans of a Catphan 504 phantom were obtained using various pitch factors (PFs). The in-plane image quality was assessed using a task-based transfer function (TTF) and a noise power spectrum (NPS). The longitudinal characteristics were evaluated using the slice sensitivity profile (SSP) and modulation transfer function (MTF). Finally, the helical artifacts and radiation doses were quantified. The in-plane TTF and NPS for IR were independent of the PF. Compared to FBP, IR provided significant noise reduction (e.g., 73.3% lower NPS at 0.2 mm⁻1) but showed slightly lower TTF for moderate-contrast objects at low frequencies. In contrast, the longitudinal resolution was directly dependent on the PF; specifically, with increasing PF, the effective slice thickness widened from 3.7 mm (Fine) to 6.0 mm (Coarse), and the MTF degraded. This was accompanied by more pronounced helical artifacts as the circularity of the object decreased from 0.95 (PF 1.0) to 0.69 (PF 3.0). The IR provided remarkable noise reduction and improved soft-tissue visualization. However, the disparate behavior of resolution between imaging directions is a key finding: while the in-plane quality is consistent across pitch factors, the longitudinal resolution degrades significantly with a higher pitch. This characteristic has direct implications for the choice of imaging protocol, particularly for high-precision treatments where geometric accuracy is paramount.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Linear Approximation in Frequency Domain
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear....
Frequency-Domain Interpretation of PD Control
The proportional control gain, combined with the...
Time and frequency -Domain Interpretation of PI Control
Acting as a low-pass filter, the PI controller slows the system's response and extends settling times. This requires...
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...