Related Experiment Video
Updated: Feb 5, 2026

Time Multiplexing Super Resolving Technique for Imaging from a Moving Platform
Published on: February 12, 2014
Extending Calibration Anchors Resolves Low-m/z Shifts in Negative-Mode Orbitrap Workflows
Francisco José Díaz-Galiano1, Bruno Le Bizec1
1Oniris, INRAE, LABERCA, 44300 Nantes, France.
High-resolution mass spectrometry calibration errors affecting low-mass ions were identified. Extending calibration with in-source fragmentation anchors resolved these systematic deviations in Q Exactive instruments.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
Background:
- Accurate mass measurements are critical for identifying fragment ions in high-resolution mass spectrometry.
- Perfluoroalkyl sulfonic acids (PFSAs) analysis on Q Exactive Orbitrap instruments showed systematic mass errors for low-mass ions (below m/z 100).
Purpose of the Study:
- To investigate the cause of systematic mass deviations observed for low-mass ions in Q Exactive Orbitrap instruments.
- To develop and validate a method for correcting these mass accuracy errors.
Main Methods:
- Comparative analysis of perfluoroalkyl sulfonic acids (PFSAs) using Q Exactive Orbitrap and Exploris 120 mass spectrometers.
- Evaluation of standard calibration solutions versus extended calibration with in-situ generated low-mass anchors.
Main Results:
- A systematic deviation of approximately -6 ppm was observed for the O3S- fragment (m/z 79.9574) on the Q Exactive, while higher m/z ions were accurate.
- The deviation was traced to the limited calibration range of the standard Q Exactive calibration solution, not an intrinsic ion property.
- Extending calibration with additional low-m/z anchors generated in-situ fully corrected the systematic errors for ions below m/z 100.
Conclusions:
- The limited calibration range of standard solutions is the source of systematic mass errors for low-m/z ions in Q Exactive instruments.
- In-situ generation of low-m/z calibration anchors effectively resolves these errors without compromising accuracy at higher m/z.
- This calibration strategy enhances the reliability of high-resolution mass spectrometry analysis for compounds with low-mass fragments.
More Related Videos
10:44One-step Negative Chromatographic Purification of Helicobacter pylori Neutrophil-activating Protein Overexpressed in Escherichia coli in Batch Mode
Published on: June 18, 2016
06:52Optimized Workflow for Iterative Bleaching Extends Multiplexity Imaging of Highly Autofluorescent Clinical Samples
Published on: July 11, 2025
Related Concept Videos
The Anchoring-and-Adjustment Heuristic
Anchoring Junctions
Lipids as Anchors
The carboxy-terminal of most of the prenylated proteins, such as Ras proteins, contains...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...