Related Experiment Video
Updated: Feb 5, 2026

Time Multiplexing Super Resolving Technique for Imaging from a Moving Platform
Published on: February 12, 2014
Extending Calibration Anchors Resolves Low-m/z Shifts in Negative-Mode Orbitrap Workflows
Francisco José Díaz-Galiano1, Bruno Le Bizec1
1Oniris, INRAE, LABERCA, 44300 Nantes, France.
Abstract:
Reliable mass accuracy is essential for the confident identification of diagnostic fragment ions in high-resolution mass spectrometry. During the analysis of perfluoroalkyl sulfonic acids (PFSAs) with a Q Exactive Orbitrap, we consistently observed the characteristic O3Ṡ- fragment, expected at a mass-to-charge ratio (m/z) of 79.9574, reported at m/z 79.9568, a systematic deviation of approximately -6 parts per million (ppm). Similar errors affected other ions below m/z 100, while precursors and higher-m/z fragments remained within the specification. Comparison with an Exploris 120 instrument, calibrated with anchors down to m/z 59, confirmed that the deviation is not intrinsic to the ion but originates from the limited calibration range of the Q Exactive standard calibration solution. We demonstrate that extending calibration to include additional low-m/z anchors generated in situ by in-source fragmentation fully corrects the error without affecting the accuracy at a higher m/z. This adjustment resolves systematic deviations for ions below m/z 100 in the Q Exactive instruments.
More Related Videos
10:44One-step Negative Chromatographic Purification of Helicobacter pylori Neutrophil-activating Protein Overexpressed in Escherichia coli in Batch Mode
Published on: June 18, 2016
06:52Optimized Workflow for Iterative Bleaching Extends Multiplexity Imaging of Highly Autofluorescent Clinical Samples
Published on: July 11, 2025
Related Concept Videos
The Anchoring-and-Adjustment Heuristic
Anchoring Junctions
Lipids as Anchors
The carboxy-terminal of most of the prenylated proteins, such as Ras proteins, contains...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...