Inter-laboratory variability in vancomycin measurement in an external quality assessment scheme 2016-2025
Anders Larsson1, Jonathan Cedernaes1, Mats B Eriksson2
1Department of Medical Sciences, Clinical Chemistry and Pharmacology, Uppsala University, Uppsala, Sweden.
Introduction:
Vancomycin is a critical antibiotic for treating severe Gram-positive infections. Due to its narrow therapeutic window, reliable plasma concentration measurements are essential for dose adjustment and avoiding toxicity. However, method-dependent analytical variability between immunoassay platforms may compromise result comparability.
Methods:
Data from 78 external quality assurance (EQA) samples distributed between 2016 and 2025 by Labquality (Helsinki, Finland) and Equalis (Uppsala, Sweden) were analyzed. Results from laboratories using immunoassays from Abbott Laboratories (n = 1391), Beckman Coulter (n = 172), Roche Diagnostics (n = 3584), Ortho Clinical Diagnostics (n = 78), Siemens Healthineers (n = 1670), Thermo Fisher (n = 152), and ARK Diagnostics (n = 33) were included (total = 7104). The mean consensus value for each sample was used as reference.
Results:
Across the study period, Abbott Laboratories (+4.7%), Thermo Fisher (+12%), Ortho Clinical Diagnostics (-4.6%), and Siemens Healthineers (-4.9%) showed systematic biases relative to the consensus mean, whereas Beckman Coulter (-0.3%), Roche Diagnostics (<0.1%), and ARK Diagnostics (-2.4%) displayed good agreement. Temporal trends indicated method-specific drifts, most notably for Beckman Coulter (-20.2%) and Siemens Healthineers (+12.5%). Coefficients of variation ranged from 0.9% (ARK Diagnostics) to 12.1% (Beckman Coulter).
Conclusion:
Considerable intermethod bias and temporal drift exist among commonly used vancomycin immunoassays, underscoring the need for improved calibration harmonization and traceability to ensure consistent therapeutic drug monitoring.
Related Concept Videos
The Z-Scheme of Electron Transport in Photosynthesis
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Random Variables
Uppercase letters such as X or Y denote a random variable. Lowercase letters like x or y denote the value of a random variable. If X is a random variable, then X is written in words, and x is given as a number.
For example, let X = the...
Graphs of Equations in Two Variables
Variables Affecting Phosphorescence and Fluorescence