Quality assurance of online adaptive radiotherapy workflows using film dosimetry in a 3D printed thorax
Daan Hoffmans1,2, Koen Nelissen1,2, Eva Versteijne1,2
1Amsterdam UMC Location Vrije Universiteit Amsterdam, Department of Radiation Oncology, De Boelelaan 1117, Amsterdam, the Netherlands.
Background And Purpose:
Quality Assurance for online adaptive radiotherapy (oART) can be challenging. Several tests can demonstrate the dosimetric and position accuracy, but commercial phantoms are often not anatomically representative. The aim of this study was to investigate the accuracy of cone-beam computed tomography guided oART palliative and breast cancer trials by using a 3D printed thorax anthropomorphic phantom.
Materials And Methods:
An anthropomorphic phantom was 3D printed for this study which accommodates film through the spine, breast, heart, and lungs. Dose was measured for spine and breast treatment plans, whilst variations were simulated which can occur during treatment. Measurements were compared to calculated dose on the planning (pCT) and synthetic computed tomography (sCT) using gamma pass rate criteria of minimal 95 % (for gamma of 4 %/2 mm). Differences between the mean gamma were tested for significance.
Results:
Measurements done with positional and target volume changes showed no significant difference between the gamma analyses for the pCT and sCT (p = 0.15), indicating a robust and safe workflow. For extreme variations, difference was found between gamma analyses for the pCT and sCT (p = 0.051). Pass rates were all >95 %, except for three measurements in which the sCT showed density errors up to 1000 Hounsfield Units.
Conclusions:
This QA approach for oART, which used film measurements in a custom 3D-printed anthropomorphic phantom was able to validate the accuracy of the oART workflow when anatomical deviations arise and could be suitable as end-to-end test in the future.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Quality Assurance
Muscles of the Thorax
The diaphragm is at the core of thoracic musculature, the primary muscle involved in breathing. This expansive, dome-shaped muscle marks the division between the thoracic and abdominal cavities. It...
Veins of Thorax
The azygos vein, positioned just right of the midline and anterior to the vertebral column, begins at the junction of the right ascending lumbar and subcostal veins, terminating in the superior vena cava. This vein drains blood from the right side of the thoracic wall, thoracic viscera, and posterior abdominal wall.
The...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality of Water
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...