Comparison of Mechanical Properties of Patient-Specific Direct 3D-Printed Aortic Valve for Simulation Trainings: A
Shokoufeh Cheheili Sobbi1,2, Anastasiia Pavlykova-Chertovska1,2, Silke Dreesen3
1Department of Cardiothoracic Surgery, Heart and Vascular Centre, Maastricht University Medical Centre, The Netherlands.
Objective:
This study aims to evaluate the mechanical properties of 3-dimensional (3D)-printed patient-specific aortic models for surgical simulation. The objective was to analyze the effects of postcuring time and color pigments on material stiffness, flexibility, toughness, and durability and to identify the most suitable material and processing conditions to enhance the biomechanical realism of 3D-printed aortae.
Methods:
Direct 3D printing was used to fabricate patient-specific aortic models. Samples were prepared with different postcuring times (5, 10, 15, and 20 min) and with or without color pigments (pink and orange). Uniaxial tensile testing was conducted to analyze the mechanical properties, including tensile strength, stiffness, flexibility, and strain at failure, using stress-strain curves.
Results:
Uncolored samples with a postcuring time of 15 min demonstrated the best mechanical properties for simulation training, offering a balance between flexibility, stiffness, and toughness. Colored samples exhibited lower tensile strength, reduced toughness, and increased stiffness as compared with uncolored samples.
Conclusions:
Material selection and postprocessing play a crucial role in the biomechanical accuracy of 3D-printed patient-specific aortic models. Uncolored samples with a 15 min postcuring time are optimal for surgical simulation. Future research should focus on refining postprocessing techniques and directly comparing 3D-printed models with human aortic tissue to improve realism and validation.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Comparative Excretory Systems
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Bioavailability Study Design: Healthy Subjects Versus Patients