Related Experiment Video
Updated: Feb 15, 2026

A Novel Biaxial Testing Apparatus for the Determination of Forming Limit under Hot Stamping Conditions
Published on: April 4, 2017
Acceptance testing of a 0.35 T MR-Linac: procedures, QA baseline, and system limitations
Mateb Al Khalifa1,2, Tianjun Ma1, Haya Aljuaid1
1Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA.
Purpose:
This study describes and evaluates the acceptance procedure for a ViewRay (VR) MRIdian 0.35T MR-Linac, emphasizing key challenges, limitations, and recommendations to enhance clinical performance and accuracy.
Methods:
A comprehensive acceptance test was conducted at a single institution, following the manufacturer's protocols and aligned with established acceptance guidelines. Specific tools and phantoms were used to assess three primary components: mechanical, dosimetric, and Magnetic Resonance Imaging (MRI).
Results:
Overall, the test outcomes satisfied the manufacturer's specifications. However, certain issues were identified: high couch attenuation at specific gantry angles (leading to their exclusion from treatment), variations in magnetic field homogeneity at different gantry angles, and discrepancies between TPS calculations and measurements for field output factors smaller than 0.83 cm × 0.83 cm.
Conclusion:
This work provides a detailed account of the acceptance testing procedure and establishes a QA baseline for 0.35T MR-Linac systems. In doing so, it also identifies key system limitations, such as couch attenuation, magnetic field inhomogeneity, and small-field output discrepancies, underscoring the need for careful gantry angle selection, field homogeneity optimization, and meticulous validation of very small fields.
Related Concept Videos
Limiting Reactant
Hypothesis: Accept or Fail to Reject?
There are two ways to indicate that the null hypothesis is not rejected. 'Accept' the null...
The Number e as a Limit
Types of Limits I
Limit Laws I
Limits at Infinity

