Effects of contrast medium injection pressure on angiographic image quality
Sara A Abosabie1,2, Tabea Kao1,2,3, Timo Alexander Auer1,3
1Department of Radiology, Charité - Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
Rationale And Objectives:
To evaluate if higher contrast medium injection pressure can improve image quality of digital subtraction angiography (DSA) in liver-directed interventions.
Materials And Methods:
Prospective single-center study including twelve patients with hepatocellular carcinoma (n = 11) or liver metastases (n = 1) undergoing intra-arterial therapies to systematically compare DSA image quality (primary endpoint) and radiation exposure (secondary endpoint) using two microcatheters with maximum application pressures 750 ("C750") and 1,200 PSI ("C1200"). Patients underwent two DSAs with both microcatheters placed in the common hepatic artery. Application pressure, contrast medium flow, volume, and dose area product were recorded. Image quality was evaluated using a customized questionnaire and quantified by calculating signal- and contrast-to-noise-ratios, and vessel-to-liver signal intensity ratio. Results were compared by paired t-test and Wilcoxon Signed-Rank test.
Results:
Image quality using C1200 (achieved 917 ± 94 PSI) was rated more favorably than C750 (731 ± 45 PSI). C1200 reached higher scores of artery visualization in 72% cases, and significantly outperformed C750 regarding tumor blush in 100% (p < 0.001). Contrast-to-noise-ratio were significantly higher in C1200 (p < 0.001) and vessel-to-liver intensity ratios were significantly lower (p = 0.001), both reflecting improved visualization and delineation of liver vessels. Signal-to-noise ratio did not differ significantly. Mean dose area product was comparable (C1200; 371.9 vs. 374.8 μGym2).
Conclusion:
The results suggest benefits of standardized CM injections for DSA using higher application pressure to enhance image contrast and tumor demarcation during IAT.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI