Inducible displacement CT for implant loosening detection: a scoping review on methods, validation, and challenges
Maaike A Ter Wee1, Johannes G G Dobbe2, Arthur J Kievit3
1Department of Orthopaedic Surgery and Sport Medicine, Amsterdam UMC, location AMC, Amsterdam; Department of Biomedical Engineering and Physics, Amsterdam UMC, location University of Amsterdam, Amsterdam; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, the Netherlands. m.a.terwee@amsterdamumc.nl.
Background And Purpose:
Inducible displacement CT (ID-CT) is an emerging method for diagnosing implant loosening by (i) acquiring CT scans under different joint loading conditions, (ii) analyzing scans via segmentation and registration, and (iii) quantifying and visualizing relative implant-bone displacement. With multiple centers approaching these steps differently, this scoping review aimed to summarize current methodologies and key challenges.
Methods:
PubMed, Cochrane, and Embase were searched for clinical and experimental ID-CT studies on spinal and arthroplasty implants. Data was extracted using a table based on updated CT-radiostereometric analysis (RSA) guidelines, including study characteristics, CT acquisition parameters, image analysis methods, validation approaches, outcomes, and loading protocols. Diagnostic studies were assessed with QUADAS-2.
Results:
22 studies were included concerning the hip (10), knee (7), spine (4), and wrist (1), covering clinical and experimental (phantom and cadaveric) designs. Loading protocols varied widely, applying compressive, rotational, and angular forces with external stabilization or loading devices. CT acquisition ranged from µCT to conventional and weight-bearing CT, with variation in kVp, mAs, reconstruction spacing, and metal artefact reduction. Image-analysis workflows were broadly similar, though reporting of algorithms and displacement metrics was often incomplete. Diagnostic validation was limited by variable reference standards, non-prespecified loosening thresholds, and non-blinded assessments, undermining true performance. Technical validation, although often precise, did not cover the full ID-CT pipeline (i.e., complete loading protocol), leaving key sources of measurement variance untested.
Conclusion:
We showed that ID-CT is used with a wide variation in approach and limited reporting prevents the establishment of its true diagnostic accuracy.