Bridging precision and practice: dual validation of a high-resolution detector array for beam profiling and patient
Sandeep Singh1,2, Supratik Sen3, Abhay Kumar Singh4,5
1Division of Medical Physics, Department of Radiation Oncology, Rajiv Gandhi Cancer Institute and Research Centre, New Delhi, 110085, India. sandeep.singh82a@gmail.com.
Abstract:
This study assessed a high-resolution ionisation chamber-based PTW 1600SRS detector array (array) for beam profile analysis and patient-specific quality assurance (PSQA) in CyberKnife (CK) stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT). The goal was to determine its suitability for small fields and non-isocentric delivery, which is unique to robotic platforms. Detector performance was examined for dose linearity, reproducibility, beam profiles, output factors, dose-rate dependence, and verification of the Iris collimator field size. Results were benchmarked against diode-based commissioning data. Clinical applicability was tested by retrospectively verifying 20 intracranial SRS and 20 extracranial SBRT plans using gamma analysis with criteria ranging from 3%/3 mm to 1%/1 mm, as well as 4%/1 mm. The detector showed strong dose linearity (R2 = 0.999) and stable reproducibility. Beam profiles matched commissioning values within 0.5 mm, and output factors agreed within 2% for most collimators, with a maximum deviation of 3% at 5 mm. Dose-rate variation remained within 2.5% across relevant SADs. Iris collimator field sizes were consistent with reference measurements. Clinical validation achieved high passing rates, all with tolerance limit of > 95%. It enables accurate beam characterization and reliable PSQA in CK treatments. This work provides the first combined evaluation of beam analysis and clinical validation for this detector on a robotic radiosurgery system, supporting its routine use in small-field quality assurance.