Related Experiment Video
Updated: Feb 28, 2026

Combining 3D-Printing and Electrospinning to Manufacture Biomimetic Heart Valve Leaflets
Published on: March 23, 2022
Fibrotic Tissue Ingrowth of Transcatheter Heart Valve Frames
Hacina J Gill1, Brooke G K MacLeod1, Noah Tregobov1
1Centre for Cardiovascular Innovation, St. Paul's and Vancouver General Hospital, Vancouver, British Columbia, Canada; Cardiovascular Translational Laboratory, Providence Research & Centre for Heart Lung Innovation, Vancouver, British Columbia, Canada; Centre for Heart Valve Innovation, St. Paul's Hospital, University of British Columbia, Vancouver, British Columbia, Canada.
Background:
Coronary access and neosinus washout are critical considerations following transcatheter aortic valve replacement (TAVR). However, the impact of transcatheter heart valve (THV) pathology on these factors remains unexplored.
Objectives:
The aim of this study was to assess THV frame obstruction due to fibrotic tissue ingrowth and its potential impact on coronary access and neosinus flow.
Methods:
Thirty-four SAPIEN platform and 10 Evolut/CoreValve (EV/CV) platform explanted THVs were evaluated for frame cell obstruction. Patient data and clinical imaging were reviewed to assess relationships with frame fibrosis. Cannulation feasibility was assessed using 6-F catheter passage. Redo-TAVR was modeled using a SAPIEN 3 device implanted at EV/CV node 5. Histologic staining was performed to analyze endothelial-like cells, activated platelets, and fibrotic remodeling.
Results:
Across all explanted THVs, 79.5% of frame exhibited fibrotic ingrowth, with implant duration correlating with obstruction severity (P = 0.003). Fibrotic ingrowth was more commonly seen in the upper and lower regions of EV/CVs and in the upper cells adjacent to commissures in SAPIEN THVs. Fibrotic ingrowth prevented cannulation with a 6-F catheter in 13.0% of cells in areas of likely cannulation to achieve coronary access. Greater fibrotic ingrowth was associated with damage to native structure at THV explantation. Redo-TAVR models had reduced accessibility, with up to 76% unable to be fenestrated by a 6-F catheter. Frame adhesions were not detectable by standard clinical imaging. Histologic analysis demonstrated progressive granulation tissue formation, which became more organized in later explantations, with a median thickness of 0.5 mm.
Conclusions:
Fibrotic THV frame ingrowth is a common phenomenon that may complicate THV explantation and impede neosinus washout and coronary access after TAVR.