Navigating preanalytical challenges: a real-world study on single-tube pneumatic tube systems
Lisha Li1, Xiaofei Zhang2, Nannan Xu1
1Department of Clinical Laboratory, The First People's Hospital of Wenling, Affiliated Wenling Hospital, Wenzhou Medical University, Taizhou, China.
Frontiers in Physiology
|February 26, 2026
Summary
Pneumatic tube systems (PTS) can transport lab samples quickly, but delays before transport may affect test results. Immediate PTS transport is suitable for biochemical analysis, but resting times impact accuracy.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Preanalytical Variables
Background:
- Pneumatic tube systems (PTS) enhance laboratory sample transport efficiency.
- Potential preanalytical errors, such as hemolysis and altered test results due to sample resting times before PTS transport, require investigation.
Purpose of the Study:
- To evaluate the impact of a novel single-tube PTS with deceleration on routine biochemical parameters.
- To assess the effects of pre-transport sample resting times on the accuracy of biochemical tests.
Main Methods:
- A prospective study involving 34 outpatients, with four blood samples collected per participant.
- Samples were allocated to immediate PTS transport, PTS after 15-min rest, PTS after 30-min rest, or manual delivery.
- Twenty-two biochemical parameters were measured, with results compared to manual delivery and assessed against allowable error limits.
Main Results:
- Lactate dehydrogenase (LDH) levels were significantly elevated in all PTS groups compared to manual delivery.
- The bias for LDH was clinically acceptable only with immediate PTS transport; delays increased bias beyond acceptable limits.
- Glucose showed significant but clinically acceptable variations; other parameters remained within acceptable limits.
Conclusions:
- The single-tube PTS is appropriate for transporting blood specimens for biochemical analysis.
- Minimizing pre-transport delays is crucial for maintaining the reliability of biochemical results from silica-clot activator tubes.
- Resting time before PTS transport should be a critical factor in laboratory validation protocols.
Keywords:
blood specimen collectionclinical chemistryhemolysislaboratoriespneumatic transportation system