Background:
A high burden of right ventricular pacing predisposes patients to pacing-induced cardiomyopathy. Cardiac resynchronization therapy with defibrillator (CRT-D) effectively addresses this by delivering biventricular pacing, and a high rate of such pacing is essential for optimal clinical response.
Case Summary:
We present a patient with high-grade atrioventricular block who developed 2:1 ventricular pacing after upgrade to CRT-D given pacing-induced cardiomyopathy. The condition was ultimately confirmed to be due to intermittent T-wave oversensing. We successfully restored atrioventricular sequential pacing by adjusting ventricular sensing sensitivity and administering antiarrhythmic medication.
Discussion:
CRT-D serves as a crucial therapeutic approach for cardiac insufficiency; however, a low biventricular pacing percentage cannot effectively improve cardiac function. Therefore, timely detection and appropriate intervention are essential to ensure adequate biventricular pacing.
Take-Home Messages:
It is important to perform optimal postprocedural management of CRT-D patients to ensure biventricular pacing. Relevant clinical data should be provided when troubleshooting T-wave oversensing.
More Related Videos
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Cardiomyopathy II: Dilated Cardiomyopathy
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias VI: Management of Dysrhythmias