A combined automatic organ at risk contouring and synthetic Computed Tomography solution for pelvic Magnetic
Jonathan J Wyatt1,2, Sandeep Kaushik3,4, Cristina Cozzini3
1Translational and Clinical Research Institute, Newcastle University, Newcastle, UK.
Background And Purpose:
Magnetic Resonance (MR)-only radiotherapy improves treatment accuracy and efficiency but requires a synthetic Computed Tomography (sCT) for dose calculations. Automatic MR-based Organs At Risk (OARs) contouring could further improve consistency and efficiency. A combined automatic MR OAR contouring and sCT solution has been developed, with each component separately validated. This study aimed to evaluate this combined MR-only workflow against a manual MR-CT workflow for pelvic cancers.
Materials And Methods:
Radiotherapy MR and CT scans were acquired for 20 patients (10 prostate, 4 rectum, 6 anus cancer). The MR-only workflow used sCTs and automatic contours for the bladder, bowel bag, femoral heads, rectum, penile bulb, prostate, seminal vesicles and urethra along with manual target contours to optimise a MR-only plan. Doses were compared to the current clinical standard of deformably registered CT and manual MR OAR contours (MR-CT).
Results:
Mean dose differences were ; 1 percentage point relative volume for all OARs except the bowel bag. Mean absolute dose differences were also for all OARs except bowel bag in ano-rectal patients and except bowel bag, bladder V , V , V and rectum V in prostate patients.
Conclusions:
The automatic MR OAR contouring and sCT workflow had similar dose differences to manual inter-observer variability for all OARs except the lower dose bladder and rectum volumes (prostate patients only) and bowel bag (prostate and ano-rectal patients). This combined OAR contouring and MR-only workflow could enable more efficient and consistent pelvic radiotherapy treatment.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Magnetic Resonance Imaging