Knowledge based automated planning for non-coplanar VMAT stereotactic radiosurgery in ocular malignancies
Chase Cochran1, Shane McCarthy1, William St Clair1
1Medical Physics Graduate Program, Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky, USA.
This study shows that RapidPlan modeling significantly reduces ocular stereotactic radiosurgery/radiotherapy planning time while improving plan quality and organ-at-risk sparing. The HARP model offers an efficient and effective approach for ocular malignancy treatment.
Area of Science:
- Radiation Oncology
- Medical Physics
- Ophthalmology
Background:
- Stereotactic radiosurgery/radiotherapy (SRS/SRT) is a less invasive treatment for ocular malignancies.
- SRS/SRT planning is complex, time-consuming, and planner-dependent.
- Improving planning efficiency, consistency, and quality is crucial.
Purpose of the Study:
- To demonstrate the efficacy of RapidPlan modeling combined with HyperArc for ocular SRS/SRT.
- To evaluate improvements in planning efficiency, consistency, and quality.
- To compare model-based planning with manual planning for ocular diseases.
Main Methods:
- Developed and trained a HyperArc-based RapidPlan (HARP) model on 80 ocular malignancy datasets.
- Compared HARP model-based plans with manually generated plans for 20 test datasets.
- Utilized Eclipse V16 with Acuros XB, TrueBeam LINAC, and Millennium 120 MLCs for planning.
- Performed EPID-based patient-specific QA and Monte Carlo (MC) second-check.
Main Results:
- HARP model-based plans showed comparable target coverage and conformality to manual plans.
- Significant improvements in organ-at-risk (OAR) sparing were observed (e.g., optic nerve, lens, lacrimal gland).
- Planning time reduced from ~120 minutes to ~15 minutes; modulation factor, monitor units, and beam-on times decreased.
- Patient-specific QA and MC second-check results improved with HARP model-based plans.
Conclusions:
- RapidPlan-generated ocular SRS plans match or exceed manual plan quality.
- Significant reduction in planning time was achieved without compromising plan quality.
- The HARP model demonstrates a promising advancement for ocular SRS/SRT planning and treatment.