Accuracy of the Withings BPM Pro 2 (WPM07) Device for Self-Blood Pressure Measurements in the General Populations
Parounak Zelveian1, Zoya Hakobyan1, Jirar Topouchian2
1Center of Preventive Cardiology, Hospital N2 CJSC, Yerevan, Armenia.
Background:
The accuracy of blood pressure (BP) measurements performed by electronic devices is crucial in clinical practice. International professional societies emphasize the need for rigorous, standardized validation of electronic BP measuring devices to ensure their accuracy and reliability.
Objective:
This study evaluated the accuracy of the Withings BPM Pro2 (WPM07) upper‑arm oscillometric BP monitor in the adult general population using the ISO 81060‑2:2018/AMD 2:2024 Universal Standard.
Methods:
Following the ISO same‑arm sequential measurement protocol, 89 eligible pre-selected participants underwent paired test‑device and mercury reference auscultatory measurements. Nine BP measurements were performed on each participant, four with the tested device and five with the auscultatory method. The accuracy of BP measurements was assessed based on the difference between simultaneous measurements taken by two observers and those displayed by the tested device using ISO Criterion 1 (mean and standard deviation of test-reference differences) and Criterion 2 (within‑subject variability of averaged difference values).
Results:
Eighty-nine participants underwent the BP measurements protocol, data from 87 were analyzed after excluding 2 participants due to excessive BP variability. Inter‑observer agreement for reference measurements was 0.64 ± 2.14 mmHg for systolic BP (SBP) and 0.39 ± 1.98 mmHg for diastolic BP (DBP). ISO criterion 1 was met with mean test-reference differences of 1.0 ± 4.9 mmHg (SBP) and 0.6 ± 4.1 mmHg (DBP) below ISO requirements (≤ 5±8 mmHg). ISO criterion 2 was also satisfied, with within‑subject standard deviations of 3.50 mmHg (SBP) and 3.01 mmHg (DBP), well below the ISO thresholds (≤6.87/6.91 mmHg).
Conclusion:
The BPM Pro2 (WPM07) fulfills the ISO 81060‑2:2018/AMD 2:2024 accuracy requirements and is suitable for reliable home BP self‑measurement in adults.
More Related Videos
05:51Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure