Improved Luminal Visualization of Flow Diverters after Deployment Using Non-Contrast-Enhanced 3D-TOF-MRA: A
Shin Hiraguchi1, Shuzo Kanasaki1, Hiroki Shibasaki1
1Department of Radiology, Takeda Hospital, Koseikai Medical Foundation, Kyoto, Japan.
Objective:
This study aimed to improve the accuracy of luminal evaluation of flow diverters (FDs) using non-contrast-enhanced 3D time-of-flight MRA (3D-TOF-MRA).
Methods:
The Flow-Redirection Endoluminal Device (FRED) and Pipeline Flex Embolization Device with Shield Technology (Pipeline) were deployed in phantom vessels with an inner diameter of 4 mm, and a phantom vessel without an FD was used as the reference. A diluted Gd contrast agent, adjusted to achieve a T1 relaxation time equivalent to that of blood, was circulated through the phantom vessels under steady flow conditions. Non-contrast-enhanced 3D-TOF-MRA images were acquired with varying echo time (TE), flip angle (FA), and Tilted Optimized Non-Saturating Excitation (TONE) ramp settings. Source images obtained from the central and edge regions of the stents were used to evaluate the relative in-stent signal (RIS) and lumen area ratio. In addition, the signal difference between the inflow and outflow regions was calculated and assessed as the saturation ratio.
Results:
RIS values for both FRED and Pipeline decreased with increasing TE at both the central and edge regions. A similar decreasing trend was observed for the area ratio. Overall, FRED demonstrated relatively higher RIS and area ratio values than Pipeline. Regarding the FA variation, the peak RIS at the center regions occurred at an FA of 35° for both FDs, followed by a plateau. In the TONE ramp analysis, a setting of 30% yielded a saturation ratio closest to 100%.
Conclusion:
These findings suggest that luminal visualization of FDs can be improved by optimizing 3D-TOF-MRA acquisition parameters, particularly the TE, FA, and TONE ramp settings.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
11:16In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...