Comparison of Ion Selective Electrodes for the Measurement of Sodium and Potassium in Human Milk
Victoria Tennert1, Kathleen F Arcaro2, Stephen J Eyles2
1University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Background:
Accessible and cost-effective measures of human milk biomarkers are needed for lactation screening and management. Sodium (Na), potassium (K), and the sodium-potassium (Na/K) ratio in human milk show promise as indicators of secretory activation and mammary gland health.
Research Aim:
This validation study compares the accuracy and precision of handheld and benchtop ion selective electrodes (ISE) as compared to inductively coupled plasma mass spectrometry (ICP-MS) for the measurement of Na and K in defatted and whole milk.
Methods:
Milk samples (N = 27) were assessed for Na and K in defatted and whole milk using ICP-MS, handheld, and benchtop ISE instruments. Bland-Altman analysis was performed to compare Na, K and Na/K measurements in whole versus defatted milk and between instruments.
Results:
All three assay methods provided high precision for Na and K concentrations in both whole and defatted milk. Na/K ratios were similar between whole and defatted milk for both ICP-MS and the handheld ISE. Measurements of Na and Na/K were systematically higher with both ISE instruments as compared to ICP-MS. However, the systematic difference was approximately three times greater in the benchtop ISE as compared to the handheld ISE. This resulted in identification of a larger number of cases of elevated mammary epithelium permeability (MEP) with the benchtop ISE as compared to handheld ISE or ICP-MS.
Conclusions:
Meaningful assessment of Na and Na/K in whole milk is possible using the handheld ISE. Its portability and ease of use suggest potential for assessing Na and K in human milk at the point-of-care for use in both research and clinical applications.
Related Concept Videos
Potentiometry: Membrane Electrodes
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Overview
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Qualitative Analysis
For instance, group IV...
Regulation of Sodium and Potassium
Sodium Regulation
Sodium ions make up approximately 90% of extracellular cations, with a normal blood plasma concentration of 136–148 mEq/L. A decrease in blood volume and pressure triggers the release of renin from granular cells in the juxtaglomerular complex (JGC), primarily...