AV nodal ablation, where are we now? Insight from real-world Northern Irish multi-centre registry data
James Mannion1, Niall Leahy2, Kathryn Hong3
1Specialist Registrar in Cardiology.
Atrioventricular nodal (AVN) ablation with pacemakers is effective for symptomatic atrial fibrillation (AF) when rhythm control fails. This study shows a safe and durable right-sided approach, mitigating concerns about pacemaker-induced cardiomyopathy.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- Atrioventricular nodal (AVN) ablation with permanent pacemaker implantation, known as 'pace and ablate', is an option for symptomatic atrial fibrillation (AF) refractory to rhythm control.
- Concerns exist regarding pacemaker dependence and potential pacemaker-induced cardiomyopathy (PICM).
- Conduction-system pacing is recognized for its potential to mitigate PICM, suggesting a future increase in AVN ablation procedures.
Purpose of the Study:
- To evaluate the safety and efficacy of a retrospective, multicentre AVN ablation approach over a one-year period.
- To assess the durability of ablation lesions and procedural outcomes.
- To compare findings with existing data from UK and European centres.
Main Methods:
- Retrospective review of patients undergoing AVN ablation over one year.
- Utilisation of radiofrequency energy for ablation.
- Staged approach with device insertion preceding ablation.
Main Results:
- Successful AVN ablation with durable lesions observed at one year.
- Median radiofrequency application time: 2.67 minutes.
- Median procedure time: 55.5 minutes; median fluoroscopy time: 3.5 minutes; median radiation dose: 11 mGy.
- No acute or subacute complications noted up to three months post-procedure.
Conclusions:
- The right-sided AVN ablation approach, performed after device insertion, yields results comparable to other centres.
- The procedure demonstrated safety and efficacy, with durable lesions at one year.
- This practice is anticipated to become more prevalent due to risks associated with physiological pacing compared to standard methods.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
09:13Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Electrophysiology of Normal Cardiac Rhythm
Dysrhythmias II: Classification of Tachyarrhythmias
Cardiopulmonary Resuscitation III: AED Use
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias